Chroma

Chroma 19501-K

Partial Discharge Tester
0 – 500 V DC Spannungsbereich
0 – 10 A Strombereich
bis zu 1500 W Ausgangsleistung
± 0,1 % FS Messgenauigkeit Spannung
± 0,2 % FS Messgenauigkeit Strom
Datenblatt anfordern
Attribut Wert
Spannungsbereich 0 – 500 V DC
Strombereich 0 – 10 A
Ausgangsleistung bis zu 1500 W
Messgenauigkeit Spannung ± 0,1 % FS
Messgenauigkeit Strom ± 0,2 % FS
Schnittstellen RS-232, GPIB, USB
Betriebstemperatur 0 °C – 40 °C
Schutzfunktionen Überspannungsschutz, Überstromschutz, Überhitzungsschutz
Versorgungsspannung 100 – 240 V AC, 50/60 Hz
Abmessungen (B x H x T) 425 mm x 88 mm x 550 mm
Chroma 19501-K Options:
  • 19501-K - 19501-K - Partial Discharge Tester
  • 19501-K - A195001 - PD Calibrator
Kalibrierungsformular Chroma 19501-K

Reparaturformular Chroma 19501-K
Maximum file size: 40MB

Keine verwandten Beiträge

Weitere Informationen
Model19501-K

 CHROMA 19501-K Partial Discharge Tester

 

The Chroma 19501-K Partial Discharge Tester is an instrument equipped with AC hipot test and partial discharge detection functions, providing 0.1kV~10kV of AC output, 0.01 µA~ 300 µA of leakage current, and 1pC~2000pC partial di scharge detection range for measurement. It is specifically designed for testing high voltage semiconductor components and high insulation materials. It  is compliant to the IEC60270-1 standard in particular the partial discharge testing requirements for high voltage. It adopts narrowband measuring technology to perform the PD tests, and displays the measured results on the screen with intuitive values (pC) allowing users to clearly understand the DUT (device under test) test result.

In addition to IEC60270-1, the Chroma 19501-K PD tester is also IEC60747-5-5 and VDE0884 compliant. Its embedded IEC60747-5-5 measuring methods can meet the production test requirements of optocouplers, providing a user-friendly operating interface.  The Chroma 19501-K is equipped with 200pC and 2000pC two ranges to measure from 1pC to 2000pC, and the best resolution in range 200pC is 0.1pC. Its high accuracy measurement and intuitive display of measured results on the screen facilitate the assessment and analysis of small discharges for highly insulated objects.  The partial discharge tester provides hipot test and partial discharge detection for high insulation withstand components such as high voltage optocouplers, high voltage relays, and high voltage switches to assure quality and reliability.

 

Specifications for the  CHROMA 19501-K Partial Discharge Tester Include:

AC Output Voltage Range 0.10 kV~10.00 kV, steps 0.01 kV
Cutoff Current 0.01 µA~300.0 µA
Frequency 50Hz, 60Hz ± 0.1%, sine wave
PD detection delay time 0~9.9 sec., steps 0.1 sec

 