The Keysight N7788B Component Analyzer uses proprietary technology to push the limits of today's optical technology. The Keysight N7788B is comparable with the well-known Jones-Matrix-Eigenanalysis (JME) which is the standard method for measuring Polarization Mode Dispersion (PMD) or differential group delay (DGD) of optical devices.
The new single scan technology from Keysight, offers a range of advantages to test a complete set of parameters including:
DGD/ PMD
PDL
Power / Loss
TE / TM-Loss
2nd-order PMD
Principal States of Polarization (PSPs)
Jones and Mueller Matrices
Features of the N7788B Optical Component Analyzer include:
High throughput: A complete analysis across the C and the L band is performed in less than 10 seconds
Software Drivers: A range of software drivers is available for external control of the system. This allows easy integration in common ERP systems.
Remote control: Control of the instrument through LAN or via the Internet is supported. This supports automation as well as trouble shooting.
Report Generation: Generating PDF reports is supported. The content including layout is configurable by the user.
Real time power readout: High throughput measurement of non-connectorized components is supported by providing a real time power readout which enables fiber coupling of the new device.
Barcode Scanner: Using Barcode scanner is supported for quick transfer of the DUT serial number.
Specifications of the Keysight N7788B include:
Wavelength (must select appropriate option at time of order):
1270 nm to 1375 nm (Opt. 300, O-Band)
1270 nm to 1375 nm, 1460 nm to 1620 nm (Opt. 400, O/C/L-Band)
1460 nm to 1620 nm (Opt. 500, C/L-Band)
Operating wavelength range: 1260 nm to 1640 nm (SOP/DOP measurements are only possible outside the specification wavelength range if the user performs a manual calibration).
Differential group delay
DGD uncertainty
Resolution 2.0 nm ± (30 fs + 3.0% * DGD)
Resolution 0.1 nm ± (30 fs + 3.0% * DGD)
DGD measurement range 0 ps to 1000 ps
PMD uncertainty ± (30 fs + 2.0% * DGD)
PMD repeatability (typical) ± 3 fs
PMD measurement range 0 ps to 300 ps
Loss
PDL uncertainty
5 C-Band ± (0.05 dB + 4% * PDL)
O-Band ± (0.10 dB + 4% * PDL)
PDL repeatability (typical) ± 0.005 dB
Insertion loss uncertainty (typical)
C/L-Band ± 0.03 dB
O-Band ± 0.07 dB
Insertion loss dynamic range (typical) > 41 dB (for higher TLS power levels, increase value accordingly)
Polarization analysis
SOP uncertainty (typical) 1.5 °
DOP uncertainty ± 2.0 %
DOP uncertainty after user calibration (typical) ± 0.5 %
Wenn Sie mit Ihrem Kauf nicht vollständig zufrieden sind, senden Sie den Artikel einfach innerhalb von 30 Tagen nach Erhalt in seinem Originalzustand an uns zurück.
Zurückgesandte Artikel müssen unbenutzt sein und in der Originalverpackung mit allen beigefügten Unterlagen zurückgeschickt werden. Nach Erhalt erstatten wir Ihnen den vollen Betrag, abzüglich der ursprünglichen Lieferkosten. Alternativ tauschen wir den Artikel gerne um, wenn Sie dies bevorzugen.
MwSt. (Mehrwertsteuer)
Die Preise verstehen sich exklusive MwSt.
Innergemeinschaftliche Transaktionen: Gemäß der EU-Mehrwertsteuerrichtlinie sind B2B-Warenverkäufe über EU-Grenzen hinweg von der MwSt. befreit.
Wenn MwSt. berechnet wird, beträgt der Satz 23 %.
Außerhalb der EU wird keine MwSt. berechnet.