The Narda ELT-400 Exposure Level Tester is an innovative exposure level meter for measuring magnetic fields in the workplace and in public spaces. The ELT-400 is designed for health and safety professionals in industry, the insurance business and the service industry. The instrument can handle practically any level measurement requested in the low and medium-frequency range, simply and precisely. It is comparable to the sound level meters that are commonly used in the assessment of noise at the workplace.
Time-consuming and easily-misinterpreted measurements with a spectrum analyzer or an oscilloscope are rendered obsolete. Detailed knowledge about the evaluation procedure, field waveform or frequency is no longer needed. The results are reliable and speed and ease of use are significantly improved over all traditional methods.
Features of the Narda ELT-400 include:
Standard-compliant measurement
Acceptance testing according to CE standard IEC / EN 62233 and IEC 62311 Generic standard
Real-time measurement with RMS and peak-detector
Direct evaluation of field exposure in comparison to major standards
Automatic exposure evaluation for various waveforms
Eliminates the overestimation occurring occasionally with FFT-based evaluation
Very wide frequency range (1 Hz to 400 kHz)
Wide measurement range (up to 80 mT, type-depending)
Isotropic 100 cm2-probe (complies with standards)
Three-axis analogue signal output
External FFT analysis of signals to determine the frequency components over three-channel analog output (one channel per spatial axis), eg using a digital oscilloscope.
Technical specifications of the Narda ELT-400 include:
Frequency Range (-3 dB), selectable: 1 Hz to 400 kHz, 10 Hz to 400 kHz, 30 Hz to 400 kHz
Antenna Type: Magnetic (B) Field
Sensor Type: Isotropic Coil 100 cm2
Damage Level RMS: 160 mT The damage level reduces linearly with increasing frequency above 77.5 Hz (1/f )
Damage Level Peak: 226 mT The damage level reduces linearly with increasing frequency above 620 Hz (1/f ) (The damage level (peak) applies for pulse duration ‰¤15.6 ms and duty cycle ‰¤1/64)
Measurement Uncertainty: 4% (50 Hz to 120 kHz)
Mounting Thread: 1/4-20UNC-2B (standard thread)
Resolution (Range: Low): 0.001%
Detection, selectable: Automatic according to Selected Standard orRMS (averaging time 1 s) or Peak Value
Interface (Remote Control and Readout): RS-232 (19200 baud, 8n1, XON/XOFF), 3-Wire, 2.5 mm Stereo Jack
Operating Temperature Range: -10 °C to +50 °C
Operating Humidity Range: <95% (30 °C) or
Weight, Typical: 2 lbs. (910 g) with probe
Dimensions, Typical: 7.1 x 3.9 x 2.2 inches (180 x 100 x 55 mm) without probe / probe 11.8 x 4.9 O inches (300 x 125 O mm)
Wenn Sie mit Ihrem Kauf nicht vollständig zufrieden sind, senden Sie den Artikel einfach innerhalb von 30 Tagen nach Erhalt in seinem Originalzustand an uns zurück.
Zurückgesandte Artikel müssen unbenutzt sein und in der Originalverpackung mit allen beigefügten Unterlagen zurückgeschickt werden. Nach Erhalt erstatten wir Ihnen den vollen Betrag, abzüglich der ursprünglichen Lieferkosten. Alternativ tauschen wir den Artikel gerne um, wenn Sie dies bevorzugen.
MwSt. (Mehrwertsteuer)
Die Preise verstehen sich exklusive MwSt.
Innergemeinschaftliche Transaktionen: Gemäß der EU-Mehrwertsteuerrichtlinie sind B2B-Warenverkäufe über EU-Grenzen hinweg von der MwSt. befreit.
Wenn MwSt. berechnet wird, beträgt der Satz 23 %.
Außerhalb der EU wird keine MwSt. berechnet.