Tektronix FCA3000

| Attribut | Wert |
|---|---|
| Frequenzbereich | DC bis 300 MHz (Kanal 1 und 2), DC bis 3 GHz (Kanal 3, optional) |
| Auflösung | Bis zu 12 Stellen Frequenzauflösung |
| Messzeit | Minimale Torzeit 100 ns für schnelle Messungen |
| Eingangsimpedanz | 1 MΩ parallel zu 20 pF (Kanal 1 und 2), 50 Ω (Kanal 3) |
| Empfindlichkeit | 25 mV eff. (typisch) bei 100 MHz |
| Zeitbasisstabilität | Interner Referenzoszillator mit ±1 ppm Genauigkeit (TCXO) |
| Messparameter | Frequenz, Periode, Impulsbreite, Tastverhältnis, Anstiegszeit, Phasendifferenz |
| Anzeigetyp | LCD-Farbdisplay mit statistischer Datenanzeige (Min, Max, Mittelwert, Standardabweichung) |
| Schnittstellen | USB, GPIB und RS-232 für externe Steuerung und Datenübertragung |
| Stromversorgung | 100–240 V AC, 50/60 Hz, automatische Spannungsanpassung |
Tektronix FCA3000 Options:
- FCA3000 - MS - Medium-stability Oven Time Base
- FCA3000 - HS - High-stability Oven Time Base
- FCA3000 - RP - Rear-panel Connectors
- FCA3000 - A0 - Power Plug, North America
- FCA3000 - A1 - Power Plug, Universal Euro
- FCA3000 - A2 - Power Plug, United Kingdom
- FCA3000 - A3 - Power Plug, Australia
- FCA3000 - A5 - Power Plug, Switzerland
- FCA3000 - A6 - Power Plug, Japan
- FCA3000 - A10 - Power Plug, China
- FCA3000 - A11 - Power Plug, India
- FCA3000 - A12 - Power Plug, Brazil
- FCA3000 - E1 - Power Plug, UK and Euro (FCA3000 and FCA3100 only)
- FCA3000 - C3 - Calibration Service 3 Years
- FCA3000 - C5 - Calibration Service 5 Years
- FCA3000 - D1 - Calibration Data Report
- FCA3000 - R5 - Repair Service 5 Years
- FCA3000 - SILV200 - Standard Warranty Extended to 5 Years (FCA3000, FCA3003, FCA3100, and FCA3103)
| Model | FCA3000 |
|---|
Tektronix FCA3000 Frequency Counter / Analyzer
The Tektronix FCA3000 Frequency Counter/Analyzer offers high-precision measurements in a compact and easy-to-use package. With industry-leading frequency and time resolution, the FCA3000 comes standard with deep internal memory and a fast data transfer rate of 250k Samples/s to internal memory. With the industry's most comprehensive analysis modes, including measurement statistics, histograms, and trend plots, Tektronix FCA3000 Frequency Counter/Analyzer is capable of quickly and accurately analyze your signal.
The Tektronix FCA3000 Frequency Counter/Analyzer offers a zero dead-time measurement technique through the USB/GPIB interfaces with a raw time-stamping function to ensure correct measurements of Allan Deviation. For correct calculation of statistical parameters, the FCA3000 comes with limit-qualifying capability. By setting limits, you can isolate one cluster in your calculation. The FCA Microwave Counter Series offers industry-best throughput, saving up to 90% on testing time. The instrument has a user-friendly interface with a multi-parameter display which shows auxiliary measurements alongside your main measurement to provide you with the results you need at a glance. Its high-precision measurements and advanced analysis functions make Tektronix FCA3000 Frequency Counter/Analyzer an ideal tool for engineers, technicians, and scientists who require accurate frequency and time measurements.
Specifications for the Tektronix FCA3000 Frequency Counter / Analyzer Include:
| Frequency Range | 0.001 Hz to 20 GHz |
| Time Resolution | 100 ps |
| Frequency resolution | 12 Digit/s |
| Voltage Resolution | 3 mV |