Fluke

Fluke 5730A Multifunction Calibrator

High-Performance Multifunction Calibrator
0 to ±1100 V DC voltage range
3.5 ppm + 2.5 µV DC voltage, best 1-year accuracy (95 %)
220 mV to 1100 V, 10 Hz to 1 MHz AC voltage range
42 ppm + 8 µV AC voltage, best 1-year accuracy (95 %)
0 to ±2.2 A; ±11 A with 5725A; ±100 A with 52120A DC current range
Request datasheet
Attribute Value
DC voltage range 0 to ±1100 V
DC voltage, best 1-year accuracy (95 %) 3.5 ppm + 2.5 µV
AC voltage range 220 mV to 1100 V, 10 Hz to 1 MHz
AC voltage, best 1-year accuracy (95 %) 42 ppm + 8 µV
DC current range 0 to ±2.2 A; ±11 A with 5725A; ±100 A with 52120A
DC current, best 1-year accuracy (95 %) 35 ppm + 7 nA
AC current range 9 µA to 2.2 A, 10 Hz to 10 kHz; 11 A with 5725A; 120 A with 52120A
AC current, best 1-year accuracy (95 %) 103 ppm + 8 nA
Resistance 0 to 100 MΩ, 18 values in ×1 and ×1.9 decades
Resistance, best 1-year accuracy (95 %) 6.5 ppm
Wideband AC voltage option 300 µV to 3.5 V, 10 Hz to 50 MHz; 30 MHz option (5730A/03), 50 MHz option (5730A/05)
Wideband, best 1-year accuracy (99 %) 0.4 % + 500 µV
Display 6.5 in touchscreen, choice of nine languages
Interfaces IEEE-488 (GPIB), RS-232, USB 2.0, Ethernet, phase lock in (BNC), phase reference out (BNC)
Amplifier support 5725A, 52120A
Warm-up time Twice the time since last warmed up, maximum 30 minutes
Settling time Less than 5 seconds for all functions and ranges except as noted
Temperature Operating 0 °C to 50 °C; calibration 15 °C to 35 °C; storage −40 °C to 75 °C
Line power 47 Hz to 63 Hz; 100, 110, 115, 120, 200, 220, 230, 240 V ±10 %
Power consumption 300 VA
Safety IEC 61010-1, CAT II, 300 V
Guard isolation 20 V
Dimensions (H × W × D) 17.8 cm (plus 1.5 cm feet) × 43.2 cm × 64.8 cm
Weight (without options) 27 kg
Calibration Form Fluke 5730A Multifunction Calibrator

Repair Form Fluke 5730A Multifunction Calibrator
Maximum file size: 40MB

No Related Posts

More Information
Model5730A

The Fluke 5730A High-Performance Multifunction Calibrator is the successor to the 5720A and the reference standard for calibrating 8.5-digit digital multimeters and other high-end electrical workload. It sources DC and AC voltage up to 1100 V, DC and AC current up to 2.2 A (extendable with the 5725A and 52120A amplifiers) and resistance up to 100 MΩ, and adds a colour touchscreen, remote operation and full compatibility with MET/CAL automated procedures.