Fluke

Fluke 700PD2

Dual Pressure Module,+/-1 PSI, +/-6.9kPa
0 to 300 psi (0 to 2068 kPa) Pressure Range
Differential Pressure Type
±0.05% full scale Accuracy
0.01 mV/V/psi Output Signal
10 VDC Excitation Voltage
Request datasheet
Attribute Value
Pressure Range 0 to 300 psi (0 to 2068 kPa)
Pressure Type Differential
Accuracy ±0.05% full scale
Output Signal 0.01 mV/V/psi
Excitation Voltage 10 VDC
Proof Pressure 600 psi (4137 kPa)
Burst Pressure 1500 psi (10342 kPa)
Process Connection 1/4 NPT male
Operating Temperature -20°C to 85°C (-4°F to 185°F)
Compatible With Fluke 700 Series Pressure Calibrators
Fluke 700PD2 Options:
  • 700PD2 - FLUKE-700PCK - Pressure Calibration Kit (power supply, interface adapter, cables, and 700PC Pressure Module Calibration software)
  • 700PD2 - FLUKE-700PMP - Hand operated Pressure Pump, provides pressures up to approximately 150 psi (1000 kPa)
Calibration Form Fluke 700PD2

Repair Form Fluke 700PD2
Maximum file size: 40MB

No Related Posts

More Information
Model700PD2

 Fluke 700PD2  Dual Pressure Module

 

The Fluke  700PD2 Dual Pressure Module allows to measure pressure with the Fluke 700 Series Process Calibrators. It displays pressure readings in any of 10 diffe pressure units specified in the calibrator setup.  The pressure module measures positive and negative pressure using an internal microprocessor. It receives operating power from and sends digital information to the calibrator.  Fluke  700PD2 Dual Pressure Module features internal temperature compensation from 0 to 50 Degree Celsius for full-accuracy performance.  The calibrators automatically time-stamp and store all  saved measurements, saving  time, and eliminating errors associated with making manual notations in the field.

Fluke  700PD2 Dual Pressure Module  has two pressure fittings and measures the difference between the applied pressure on the high fitting versus the low fitting. It functions like a gauge module when the low fitting is open. For best results, it is recommended that the module be pressurized to full scale and then vented to zero pressure (atmosphere) prior to zeroing and making measurements. This module can be calibrated locally using the Fluke 700PCK Pressure Calibration Kit, a PC, and a precision pressure source.  Rugged urethane molded case of Fluke  700PD2 Dual Pressure Module protects the module from rough handling and harsh conditions.

 

Specifications for the  Fluke  700PD2 Gauge Pressure Module Include:

Resolution 0.0001 / 0.07
Measurement range 1  psi  / 6900 Pa
Reference Uncertainty (23 ± 3 °C) 0.15
Total Uncertainty 1 year 0.2

 