Chroma

Chroma 11210

Battery Cell Insulation Tester
0 – 300 V CA Plage de tension de sortie
500 VA Puissance nominale
45 Hz – 500 Hz Fréquence de sortie
0,1 V Résolution de tension
0,1 Hz Résolution de fréquence
Demander la fiche technique
Attribut Valeur
Plage de tension de sortie 0 – 300 V CA
Puissance nominale 500 VA
Fréquence de sortie 45 Hz – 500 Hz
Résolution de tension 0,1 V
Résolution de fréquence 0,1 Hz
Distorsion harmonique totale (THD) ≤ 0,5 %
Interface de communication RS-232 / GPIB (option)
Tension d'entrée 115 V / 230 V CA, monophasé
Dimensions (L x H x P) 213 mm x 88 mm x 430 mm
Poids 9 kg
Chroma 11210 Options:
  • 11210 - A112100 - Partial Discharge Detection Card
  • 11210 - A112101 - Partial Discharge Analyzer Card
  • 11210 - A112102 - PD Test Checking Kit
Formulaire d'étalonnage Chroma 11210

Formulaire de réparation Chroma 11210
Maximum file size: 40MB

No Related Posts

Plus d'informations
Model11210
Chroma 11210 Battery Cell Insulation Tester  The Chroma 11210 battery cell insulation tester is an instrument used for accurately measuring leakage cur (LC) and insulation resistance (IR) of battery jelly-roll/dry-cell as well as various insulation materials. In addition to standard LC/IR measurement, it has a unique function that detects partial discharge (PD) or flashover that may have occurred inside the insulation material during the high voltage insulation testing process. Due to special circuitry used in the Chroma 11210, no extra time is required for the PD detection function. Both LC/IR measurements and PD detection tests can be performed simultaneously and both test results will be reported automatically.With an ample charging cur capability, the Chroma 11210 can easily achieve high speed testing with automatic execution of the test sequence. This sequence can be automatically executed in 20ms or less for a single DUT (Device Under Test); meaning tests performed in the production line can achieve process rates of 50 pcs/sec or more. The contact check function in the Chroma 11210 employs advanced technology to instantly detect those which do not make secure contacts during the tests. It takes only 5ms to do a thorough contact check before and/or after the measurement. Specifications for the Chroma 11210 Battery Cell Insulation Tester Include:
  • Test voltage : up to 1KV(dc)
  • Charge cur : 50mA max.
  • Wide range of Leakage Cur (LC) measurement (10pA ~ 20mA)
  • Partial discharge/flashover detection for inspection on potential internal short circuits (option of A112100)
  • PD level and number of occurrence display
  • Built-in reliable contact check
  • Automatic test with sequence : charge-dwell-measure-discharge
  • High speed testing (20ms/device)
  •       