Hipotronics

Hipotronics H815PL-A

DC Hipot Tester, 15kV
15 kV CA Tension de sortie maximale
5 kVA Puissance de sortie
50/60 Hz Fréquence de fonctionnement
115/230 V CA Tension d'alimentation
333 mA Courant de court-circuit
Demander la fiche technique
Attribut Valeur
Tension de sortie maximale 15 kV CA
Puissance de sortie 5 kVA
Fréquence de fonctionnement 50/60 Hz
Tension d'alimentation 115/230 V CA
Courant de court-circuit 333 mA
Type de test Test diélectrique haute tension (Hipot)
Précision de mesure ±2% de la pleine échelle
Affichage Mètre analogique pour tension et courant
Protection intégrée Disjoncteur de surcharge et détection de défaut à la terre
Conformité aux normes IEEE, ANSI, IEC
Hipotronics H815PL-A Options:
  • H815PL-A - 815PL-A - 15 kV, 5 mA DC Hipot, 120 V 50/60 Hz input
  • H815PL-A - 815PL-B - 15 kV, 5 mA DC Hipot, 230 V 50/60 Hz input
  • H815PL-A - SPK1-815PL - Spare Parts Kit for 815PL
Formulaire d'étalonnage Hipotronics H815PL-A

Formulaire de réparation Hipotronics H815PL-A
Maximum file size: 40MB

No Related Posts

Plus d'informations
ModelH815PL-A

 Hipotronics  H815PL-A DC Hipot Tester

 

The  Hipotronics  H815PL-A DC Hipot Tester  is an economical solution to DC field testing of cables, terminations, motors, generators and other electrical apparatus. The Hipotronics H815PL-A and is self-contained in a single, rugged and durable enclosure, which is complete from input line cord to high voltage output cable. Cable storage space is conveniently located inside.  Hipotronics  H815PL-A DC Hipot Tester is compact and lightweight in design which makes it suitable for field orientated applications.  

The  Hipotronics H815PL-A DC Hipot Tester features accurate kV meter readings regardless of load current. Accurate voltage and cur measurements are assured via voltage readings that are taken directly at the output of the high voltage transformer and cur measurements are taken in the return leg. Safe discharging of both the test object and the high voltage transformer occur whenever the high voltage is turned off. Output power is provided by means of a full-wave voltage doubling rectifier circuit with silicon rectifiers.

 

Specifications for the  Hipotronics  H815PL-A DC Hipot Tester Include:

Output Voltage 0 -15 kV
Output  Current 5 mA
Metering Accuracy ±2% of full scale
Ripple 0.5% per mA with rated resistance

 

Clicky Copyright © 2013-present Magento, Inc. All rights reserved.