Test Equity 1016H

| Attribut | Valeur |
|---|---|
| Plage de température | -15°C à +35°C |
| Précision de température | ±0,5°C |
| Capacité de volume intérieur | 1016 litres |
| Alimentation électrique | 208-230 V, 60 Hz, monophasé |
| Consommation électrique | 5,0 kW |
| Système de régulation | Contrôleur PID numérique |
| Uniformité de température | ±1,0°C |
| Dimensions extérieures (L x P x H) | 1270 mm x 813 mm x 1981 mm |
| Poids net | approximately 454 kg |
| Conformité et certifications | UL, CE |
| Model | 1016H |
|---|
TEST EQUITY 1016H Temperature and Humidity Chamber
TestEquity Model 1016H Temperature/Humidity Chamber provides 16 Cubic Feet of chamber workspace. It is designed to fit the industry-standard 16 Cu Ft footprint, allowing you to fit a TestEquity 1016H in the same space as 16 Cu Ft models from other manufacturers. The F4T controller features a touch-screen interface, it's exceptionally easy and intuitive to use. Includes RS-232 and Ethernet interfaces. GPIB is available as an option. Provides independent protection against excess temperatures. Both high and low limits can be set. The upper LED readout indicates the actual chamber temperature while the lower readout indicates the status The limit controller will shut down the chamber and trigger the audible alarm in the event of an out of limit condition.
Unlike "atomizing" systems, the Test Equity 1016H has a heated vapor generator that has no pump to fail or orifice to clog. The stainless-steel vapor generator assembly is protected against low-water and over-temperature conditions. The immersion heater has an embedded thermocouple for fast-responding thermal protection. The optional W100 water recirculation system provides a reliable source of demineralized water for the humidity system, using ordinary tap or pre-filtered water. Non-CFC refrigerants are used in a cascade (two compressors) configuration. Expansion valves provide optimum performance throughout the operating range. Pressure regulators keep the operating pressures within safe limits regardless of load conditions. Liquid injection ensures cool compressor operation during high-temperature pull down for long life.
Specifications for the TEST EQUITY 1016H Temperature and Humidity Chamber Include:
| Temperature Range | -73 °C to +175 °C |
| Humidity Range | 10% to 95% |
| Display Resolution | 0.1% RH |
| Inside Dimensions | 30" W x 30" H x 30" D (15.6 cubic feet) |
| Outside Dimensions | 38" W x 78.5" H x 56" D |