Tektronix TDP7704

| Attributo | Valore |
|---|---|
| Larghezza di banda | 4 GHz |
| Tensione di modo comune massima | ±15 V (DC + picco AC) |
| Rapporto di reiezione di modo comune (CMRR) | 30 dB a 1 GHz |
| Impedenza di ingresso differenziale | 100 kΩ in parallelo con 0,8 pF |
| Fattore di attenuazione | 10X / 100X |
| Tensione di ingresso massima differenziale | ±8 V (10X), ±80 V (100X) |
| Tempo di salita | 120 ps |
| Compatibilità oscilloscopio | Oscilloscopi Tektronix serie DPO/DSA/MSO70000 |
| Connettore di uscita | Connettore SMA |
| Temperatura operativa | 0°C - 50°C |
Tektronix TDP7704 Options:
- TDP7704 - C3 - Calibration Service 3 Years
- TDP7704 - C5 - Calibration Service 5 Years
- TDP7704 - D3 - Calibration Data Report 3 Years (with Opt. C3)
- TDP7704 - D5 - Calibration Data Report 5 Years (with Opt. C5)
- TDP7704 - G3 - Complete Care 3 Years (includes loaner, scheduled calibration, and more)
- TDP7704 - G5 - Complete Care 5 Years (includes loaner, scheduled calibration, and more)
- TDP7704 - R3 - Repair Service 3 Years (including warranty)
- TDP7704 - R3DW - Repair Service Coverage 3 Years (includes product warranty period). 3-year period starts at time of instrument purchase
- TDP7704 - R5 - Repair Service 5 Years (including warranty)
- TDP7704 - R5DW - Repair Service Coverage 5 Years (includes product warranty period). 5-year period starts at time of instrument purchase
| Model | TDP7704 |
|---|
TEKTRONIX TDP7704 TriMode Probe
The TDP7704 TriMode Probe introduce the TekFlex connector technology, a groundbreaking advancement in connectivity and functionality. This innovative connector seamlessly integrates a high-speed signal path, power supply, and communication support for an active buffer tip, all within a single, user-friendly accessory connector. With its pinch-to-open design, attaching an accessory tip becomes effortless, requiring minimal force. When closed, the TekFlex connector establishes a secure connection, eliminating the risk of accidental disconnections. This cutting-edge connector empowers the TDP7704 probe to provide active probe tips positioned just millimeters away from the input connections, resulting in a short signal path. This configuration delivers exceptional fidelity, high impedance input, and effectively reduces signal loss, capacitance, and additive noise, ensuring precise and reliable measurements.
Specifications for the TEKTRONIX TDP7704 TriMode Probe Include:
| Bandwidth | 4 GHz |
| Rise Time (10% - 90%) | < 100 ps |
| Operating Voltage | ±4 V (2.92 tip), ±5.25 V (active tip), ±10 V (browser tip) |
| Accuracy | ±2 % |