The Keithley 2303 Power Supply provides voltage control and power monitoring for automated testing of portable, battery-operated devices. The Keithley 2303 delivers up to 5A to handle portable device loads easily, while 100nA resolution enables precise monitoring of sleep mode currents with 0.2% basic accuracy. Fast transient response simulates battery behavior, with output returning to within 100mV of the set voltage in 40 µs or less. The ability to sink cur enables the Keithley 2303 Power Supply to simulate the characteristics of a discharged, rechargeable battery for testing of battery chargers and charger-related circuitry. In addition to DC cur measurements, it can measure cur pulses as short as 60 µs and as long as 833ms.
Features and Specifications of the Keithley 2303 Power Supply include:
DC Voltage Output
Output Voltage : 0 to +15V DC
Output Accuracy : ±(0.05% + 10mV)
Programming Resolution : 5mV
Readback Accuracy: ±(0.05% + 3mV)
Readback Resolution : 1mV
Output Voltage Settling Time: 5ms to within stated accuracy
Load regulation : 0.01% + 2mV
Line Regulation : 0.5mV
Stability: 0.01% + 0.5mV
Transient Response To 1000% Load Change:
Transient Recovery Time:
<40 µs to within 100mV of previous level
<80 µs to within 20mV of previous level
Transient Voltage Drop:
<100mV, typical
<200mV, typical
Ripple and Noise (20Hz to 20MHz): 3mV rms/8mV p-p, typical
DC Current
Output Cur : 0 €“9V: 5A max >9V €“15V: 3A max. (not intended to be operated in parallel)
Source Compliance Accuracy : ±(0.16% + 5mA)
Programmed Source Compliance Resolution : 1.25mA
Readback Accuracy:
5A range: ±(0.2% + 400 µA)
5mA range: ±(0.2% + 1 µA)
Readback Resolution
5A range: 100 µA
5mA range: 0.1 µA
Cur Sink Capacity : 0 €“5V: 2A max 5V €“15V: Derate 0.1A per volt above 5V
Load regulation : 0.01% + 1mA
Line Regulation : 0.5mA
Stability: 0.01% + 50 µA
Digital Voltmeter Input
Input Voltage Range: 0 to +20V DC
Input Impedance: 1010O typical
Maximum Voltage (either input terminal) with respect to output low: €“3V, +22V
Reading Accuracy: ±(0.05% + 3mV)
Reading Resolution : 1mV
DC General
Measurement Time Choices: 0.01 to 10 PLC, in 0.01PLC steps
Average Readings: 1 to 10
Reading Time: 31ms, typical
Pulse Cur Measurement Operation
Trigger Level: 5mA to 5A, in 5mA steps
Trigger Delay: 0 to 100ms, in 10 µs steps
Internal Trigger Delay: 25 µs
High /Low /Average Mode:
Measurement Aperture Settings: 33.3 µs to 833ms, in 33.3 µs steps
Average Readings: 1 to 100
Burst Mode:
Measurement Aperture: 33.3 µs
Conversion Rate: 3600/second, typical
Number of Samples: 1 to 5000
Transfer Samples Across IEEE Bus in Binary Mode: 4800 bytes/second, typical
Long Integration Mode:
Measurement Time: 850ms (840ms) to 60 seconds in 16.7ms (20ms) steps