Keysight (Agilent)
Keysight (Agilent) 83712B
Synthesized CW Generator, .01 - 20 GHz
10 MHz to 20 GHz Frequency Range
1 kHz Frequency Resolution
< -50 dBc at < +6 dBm Harmonics
< -60 dBc at ≥ 3 kHz offset Non-Harmonics
< 50 ms to within 1 kHz Switching Speed

| Attribute | Value |
|---|---|
| Frequency Range | 10 MHz to 20 GHz |
| Frequency Resolution | 1 kHz |
| Harmonics | < -50 dBc at < +6 dBm |
| Non-Harmonics | < -60 dBc at ≥ 3 kHz offset |
| Switching Speed | < 50 ms to within 1 kHz |
| Output Power | -4 dBm to +13 dBm below 1 GHz, +11 dBm to 18 GHz, +10 dBm to 20 GHz |
| SSB Phase Noise | -76 dBc/Hz at 10 to 20 GHz, 10 kHz offset |
| Level Accuracy | ±1.0 dB at 50 MHz to 20 GHz, -4 dBm to max. (spec.) |
| Timebase Aging | 10 MHz, < 1 × 10⁻⁸ per day |
| RF Output Connector | Type-N |
| Interfaces | GPIB |
| Dimensions | 426 × 133 × 498 mm |
| Weight | 16 kg |
Keysight (Agilent) 83712B Options:
- 83712B - 1E1 - Adds 110 dB output step attenuator
- 83712B - 1E5 - Adds high-stability timebase
- 83712B - 1E8 - 1 Hz frequency resolution
- 83712B - 1E9 - 3.5 mm RF output connector
- 83712B - 0B2 - Extra operating manual
- 83712B - 0BV - Service documentation, component level
- 83712B - 0BW - Service documentation, assembly level
- 83712B - 1CM - Rack mount kit (Part number 5062-3977)
- 83712B - 1CP - Rack mount and handle kit (Part number 5062-3983)
- 83712B - 1CR - Rack slide kit (Part number 1494-0059)
The Agilent HP 83712B Synthesized CW Generator set a high standard for performance for an affordable price. The HP 83712B Signal Generator features fundamental oscillators and switched low-pass filters which deliver +14 dBm), while spectral purity is maintained even at high power levels (typical output power at frequencies below 1 GHz is +20 dBm).
Features of the HP 83712B Signal Generator include:
- Frequency range: .01 to 20 GHz
- High stability time base for frequency accuracy
- Low SSB phase noise for out-of-channel tests
- High power output for driving the LO port of mixers
- Low output power with option 1E1 for receiver sensitivity and SINAD measurements
Options and configurations
- Frequency Resolution: 1 kHz (1 Hz with Opt. 1E8)
- Output Power: -4 dBm (-110 dBm with Opt. 1E1) to +13 dBm below 1 GHz, +11 dBm to 18 GHz, +10 dBm to 20 GHz
- Timebase Aging: 10 MHz, < 1 × 10⁻⁸ per day (< 1.5 × 10⁻⁹ per day with Opt. 1E5)
- RF Output Connector: Type-N (3.5 mm with Opt. 1E9)