The Keysight N7788B Component Analyzer uses proprietary technology to push the limits of today's optical technology. The Keysight N7788B is comparable with the well-known Jones-Matrix-Eigenanalysis (JME) which is the standard method for measuring Polarization Mode Dispersion (PMD) or differential group delay (DGD) of optical devices.
The new single scan technology from Keysight, offers a range of advantages to test a complete set of parameters including:
DGD/ PMD
PDL
Power / Loss
TE / TM-Loss
2nd-order PMD
Principal States of Polarization (PSPs)
Jones and Mueller Matrices
Features of the N7788B Optical Component Analyzer include:
High throughput: A complete analysis across the C and the L band is performed in less than 10 seconds
Software Drivers: A range of software drivers is available for external control of the system. This allows easy integration in common ERP systems.
Remote control: Control of the instrument through LAN or via the Internet is supported. This supports automation as well as trouble shooting.
Report Generation: Generating PDF reports is supported. The content including layout is configurable by the user.
Real time power readout: High throughput measurement of non-connectorized components is supported by providing a real time power readout which enables fiber coupling of the new device.
Barcode Scanner: Using Barcode scanner is supported for quick transfer of the DUT serial number.
Specifications of the Keysight N7788B include:
Wavelength (must select appropriate option at time of order):
1270 nm to 1375 nm (Opt. 300, O-Band)
1270 nm to 1375 nm, 1460 nm to 1620 nm (Opt. 400, O/C/L-Band)
1460 nm to 1620 nm (Opt. 500, C/L-Band)
Operating wavelength range: 1260 nm to 1640 nm (SOP/DOP measurements are only possible outside the specification wavelength range if the user performs a manual calibration).
Differential group delay
DGD uncertainty
Resolution 2.0 nm ± (30 fs + 3.0% * DGD)
Resolution 0.1 nm ± (30 fs + 3.0% * DGD)
DGD measurement range 0 ps to 1000 ps
PMD uncertainty ± (30 fs + 2.0% * DGD)
PMD repeatability (typical) ± 3 fs
PMD measurement range 0 ps to 300 ps
Loss
PDL uncertainty
5 C-Band ± (0.05 dB + 4% * PDL)
O-Band ± (0.10 dB + 4% * PDL)
PDL repeatability (typical) ± 0.005 dB
Insertion loss uncertainty (typical)
C/L-Band ± 0.03 dB
O-Band ± 0.07 dB
Insertion loss dynamic range (typical) > 41 dB (for higher TLS power levels, increase value accordingly)
Polarization analysis
SOP uncertainty (typical) 1.5 °
DOP uncertainty ± 2.0 %
DOP uncertainty after user calibration (typical) ± 0.5 %