| Producent | ETS Lindgren |
| Model | HI-4433-STE / HI-443 |
| Produkt | Electric Field Probe / Magnetic Field Probe Kit |
| Typ | Uzywany |
The Holaday ETS-Lindgren HI-4433-STE Electric Field Probe features:
Key Features
Description
This Automatic Continuous Zeroing feature removes the need for internal zero adjustment and eliminates temperature drift errors associated with other instrumentation. The Automatic Zero feature allows the meter to be used immediately upon power-up, and eliminates concerns about finding a zero.
Standard Configuration
Options
Electrical Specifications
|
Frequency Minimum |
500 kHz |
|
Frequency Maximum |
5 GHz |
|
Battery |
20 Hours Continuous Operations |
|
Calibration Accuracy |
absolute accuracy +/- 1.0 dB |
|
Data Sample Rate |
1 - 24 Samples per Second |
|
Detection |
Isotropic |
|
Dynamic Range |
30 - 3000 V/m |
|
Isotropic Deviation |
+/- 0.5 dB |
|
Linearity |
+/- 0.5 dB full scale |
|
Response Time |
86 msec (min) |
The Holaday ETS-Lindgren HI-4433-LFH Magnetic Field Probe features:
Key Features
Description
This Automatic Continuous Zeroing feature removes the need for internal zero adjustment and eliminates temperature drift errors associated with other instrumentation. The Automatic Zero feature allows the meter to be used immediately upon power-up, and eliminates concerns about finding a zero.
Standard Configuration
Options
Electrical Specifications
|
Frequency Minimum |
300 kHz |
|
Frequency Maximum |
30 MHz |
|
Battery |
20 Hours Continuous Operations |
|
Calibration Accuracy |
absolute accuracy +/- 1.0 dB |
|
Data Sample Rate |
1 - 24 Samples per Second |
|
Detection |
Isotropic |
|
Dynamic Range |
0.3 - 30 A/m |
|
Isotropic Deviation |
+/- 0.5 dB |
|
Linearity |
+/- 0.5 dB full scale |
|
Response Time |
86 msec (min) |
The Holaday ETS-Lindgren HI-4416 EMF Readout Unit features:
Key Features
Description
The optical isolation minimizes field perturbations, allows for highly accurate readings and provides free positioning of the probe without hindering the user interface.
The instrument has on-board recording features enabling accurate and fast data collection. Up to 150 field readings can be stored for review. Other features include continual peak (max) value storage and display, units selection, (auto) range selection, individual axis field readings and bar graph for "hot spot" determinations.
Standard Configuration
Options
Specifications
|
Battery |
NiCad, approx. 80 hrs. on full charge |
|
Data Logging |
115 point scratch-pad datalogging |
|
Recorder Output |
0 - 1 VDC proportional to full scale reading |
|
Units |
V/m, V2/m2, A/m, A2/m2, mW/cm2 (Depending on probe) |
Zwroty
Jeśli nie jesteś całkowicie zadowolony z zakupu, po prostu zwróć produkt w oryginalnym stanie w ciągu 30 dni od daty otrzymania. Zwracane przedmioty muszą być nieużywane i znajdować się w oryginalnym opakowaniu wraz z dołączoną dokumentacją. Po otrzymaniu dokonamy pełnego zwrotu kosztów, z wyłączeniem pierwotnych kosztów dostawy. Alternatywnie możemy wymienić produkt, jeśli wolisz.