Fluke 6331

| Atrybut | Wartość |
|---|---|
| Zakres częstotliwości | 0,1 Hz – 1050 MHz |
| Impedancja wyjściowa | 50 Ω |
| Poziom wyjściowy RF | -137 dBm do +13 dBm |
| Rozdzielczość częstotliwości | 0,1 Hz |
| Modulacja AM | 0% – 99,9%, częstotliwość modulacji 0,1 Hz – 100 kHz |
| Modulacja FM | Odchylenie do 100 kHz, częstotliwość modulacji 0,1 Hz – 100 kHz |
| Modulacja PM (fazowa) | Odchylenie do ±10 rad, częstotliwość modulacji 0,1 Hz – 100 kHz |
| Zniekształcenia harmoniczne | Mniej niż -30 dBc przy poziomie wyjściowym 0 dBm |
| Zasilanie | 90 – 264 V AC, 47 – 63 Hz |
| Wymiary (szer. x wys. x głęb.) | 425 mm x 133 mm x 497 mm |
Fluke 6331 Options:
- 6331 - 2001-IEEE - IEEE-488 Interface, 6330/7320/7340
| Model | 6331 |
|---|
Fluke 6331 Deep-Well Compact Bath 35C to 300C
The Fluke 6331 Deep-Well Compact Bath is used for temperature calibration and best known for it's excellent temperature control, stability and uniformity. lt includes a 457 mm (18-inch) deep tank to accommodate long-stem PRTs, SPRTs, and liquid-in-glass (LIG) thermometers. Despite of it's depth, it has a relatively small footprint, minimizing fluid costs and allowing for quick temperature changes. Fluke 6331 Deep-Well Compact Bath is ideal for calibration laboratories and other environments where precise temperature control and calibration are essential.
Fluke 6331 Deep-Well Compact Bath provides stability and uniformity from ±0.007 °C to ±0.025 °C over its range from 40 °C to 300 °C. Fluke Calibration's control system automatically adds refrigeration when you need to cool down quickly, and shuts down refrigeration when you need to heat up quickly for efficient temperature adjustments. Fluke 6331 can be equipped with the optional LIG Thermometer Calibration Kit for easy and accurate calibration of LIG thermometers. Fluke 6331 Deep-Well Compact Bath is compatible with Fluke Calibration's MET/TEMP II temperature calibration software, for automated probe calibrations.
Specifications for the Fluke 6331 Deep-Well Compact Bath 35C to 300C Include:
| Temperature range | 35 °C to 300 °C |
| Stability | ±0.015 °C at 300 °C (oil 5017) |
| Uniformity | ±0.025 °C at 300 °C (oil 5017) |
| Stabilization Time | 15 €“20 minutes |