Keysight (Agilent) N432A

| Atrybut | Wartość |
|---|---|
| Zakres częstotliwości | 100 kHz – 18 GHz (zależnie od użytego czujnika mocy) |
| Zakres pomiarowy mocy | -70 dBm do +44 dBm (zależnie od czujnika) |
| Dokładność kalibracji | ±0,9% (przy 50 MHz, 1 mW) |
| Moc wyjściowa kalibratora | 1 mW (0 dBm) przy 50 MHz |
| Liczba kanałów pomiarowych | 1 kanał |
| Interfejsy komunikacyjne | GPIB, USB, LAN (LXI Class C) |
| Wyświetlacz | Kolorowy wyświetlacz LCD |
| Czas pomiaru | Minimalny czas uśredniania: 20 ms |
| Zasilanie | 100–240 V AC, 50/60 Hz |
| Kompatybilne czujniki mocy | Czujniki termistorowe serii 478A oraz 8478B (złącze czujnika termistorowego) |
Keysight (Agilent) N432A Options:
- N432A - 478A-H63 - Frequency range 100 kHz to 1 GHz, max SWR 1.8 to 300 kHz, 1.3 to 1 GHz
- N432A - 478A-H75 - Frequency range 1 MHz to 1 GHz, max SWR 1.3 except 1.05 at 50 MHz
- N432A - 478A-H76 - Frequency range 1 MHz to 1 GHz, max SWR 1.3 except 1.05 at 50 MHz. Standard lab calibration at 50 MHz
- N432A - N4998A - Thermistor sensor adaptor cable 1.5 m (5 ft)
- N432A - N4998B - Thermistor sensor adaptor cable 3 m (10 ft)
- N432A - N4998C - Thermistor sensor adaptor cable 6.1 m (20 ft)
- N432A - 908 - Rackmount kit with 1 unit with blank filler included
- N432A - 909 - Rackmount kit with 2 units side-by-side
- N432A - 0BK - English language user guide and English programming guide
- N432A - ABJ - Japanese user guide and English programming guide
- N432A - 0BF - English language programming guide
- N432A - A6J - Certificate of compliance calibration - ANSI/NCSL Z540
- N432A - 1A7 - Compliant calibration test data - ISO17025
| Model | N432A |
|---|
AGILENT N432A
The Agilent N432A Power Meteris the replacement for Agilent's legacy 432A analog thermistor power meter. The single channel, average RF power meter operates with the Agilent 8478B and 478A thermistor mounts. The N432A and thermistor mount pair is perfect for applications that require high measurement accuracy, particularly in metrology and calibration lab environments. The enhanced thermistor power meter comes with a digital color LCD display, and user-friendly front panel interface. Its built-in calibration factor table provides a more convenient and accurate method of storing calibration factors as compared to the old knob-turning approach of the 432A. The Agilent N432A is LXI-C compliant, and programmable via GPIB, USB and LAN.
Features and specifications of the Agilent N432A Power Meter include:
- Frequency range: 100 kHz to 18 GHz (sensor dependent)
- Dynamic range: -30 dBm to +10 dBm
- High Accuracy ( ‰¤ 0.1% ± 0.5 uW), excellent for 1 mW transfer calibration (with 478A-H75/H76)
- Built-in 6.5 digits ADC eliminates the need of external DMM.
- Selectable bridge resistance (100/200/300/400 Ω)
- Remote interface flexibility (GPIB, USB, LXI-C)
- Digital color LCD display