Tektronix TDP1000

Sonda różnicowa wysokiego napięcia, 1 GHz

Producent:
Model: Tektronix TDP1000
Opis: Sonda różnicowa wysokiego napięcia, 1 GHz
Arkusz danych PDF: Tektronix TDP1000

Tektronix TDP1000 Diffeial Probe, 1 GHzThe Tektronix TDP1000 High-voltage Diffeial Probe is a versatile and easy to use probe that provides high-speed electrical and mechanical performance needed for Switch Mode Power Supply, CAN/LIN Bus, and for high-speed digital system designs. The Tektronix TDP1000 Probe is designed specifically for use with, and direct connection to, Tektronix Oscilloscopes with either the TekVPI™ probe interface, or TekProbe BNC Interface.Features and Specifications of the TDP1000 Diffeial Probe include:

  • Features:
    • Versatile DUT Connectivity:
      • Small compact probe head for probing small geometry circuit elements
      • Straight pin, square pin, solder down, variable pitch standard accessories
      • Robust design for reliability
    • Provides high-speed signal acquisition and measurement fidelity with its:
      • Outstanding electrical performance
      • Versatile device-under-test connectivity
      • Ease of use
    • Connects directly to Oscilloscope with the TekVPI™ probe interface
    • Provides automatic units scaling and readout on the Oscilloscope display
    • Embeds the probe characteristics within the Oscilloscope
    • Easy access to scope-displayed probe menu:
      • for probe setup control
      • for operating status information
    • Remote GPIB/USB probe control through the Oscilloscope
    • AutoZero:  Zeros out output offset
    • Applications:
      • High-speed switch mode power supply design
      • CAN/LIN bus design
      • High-speed digital design
      • Digital design and characterization
      • Manufacturing engineering and test
      • Research and development
  • Specifications:
    • Probe Bandwidth:  1 GHz
    • Probe Attenuation:  5X, 50X
    • Probe Rise Time:  ≤350 ps
    • Diffeial Input Capacitance: 
    • Diffeial Input Resistance:  1 MΩ
    • CMRR:
      • >55 dB at 30 kHz
      • >50 dB at 1 MHz
      • >18 dB at 250 MHz (warranted at 50X attenuation)
    • Sensitivity /Noise Level:
      • 2 mVRMS (5X attenuation)
      • 1 mVRMS (50X attenuation) referred to the probe output
      • 10 mVRMS (5X attenuation)
      • 50 mVRMS (50X attenuation) referred to the probe input
    • Diffeial Mode Input Voltage:  ±42 V (DC + pk AC); 30 VRMS
    • Common Mode Input Voltage:  ±35 V (DC + pk AC); 25 VRMS
    • Input Offset Range:  ±42 V (5X or 50X attenuation)
    • Max Input Voltage (Nondestruct):  ±100 V (DC + pk AC)
    • Selectable Bandwidth Filter Limits:  100 Hz, 10 kHz, 1 MHz, Full
    • DC Reject:
      • 0.4 Hz (5X)
      • 4 Hz (50X)
    • Gain Accuracy at DC:  ±2%
    • Propagation Delay:  6.5 ns
    • Probe Interface:  TekVPI™ Probe Interface
  • Versatile DUT Connectivity:
    • Small compact probe head for probing small geometry circuit elements
    • Straight pin, square pin, solder down, variable pitch standard accessories
    • Robust design for reliability
  • Provides high-speed signal acquisition and measurement fidelity with its:
    • Outstanding electrical performance
    • Versatile device-under-test connectivity
    • Ease of use
  • Connects directly to Oscilloscope with the TekVPI™ probe interface
  • Provides automatic units scaling and readout on the Oscilloscope display
  • Embeds the probe characteristics within the Oscilloscope
  • Easy access to scope-displayed probe menu:
    • for probe setup control
    • for operating status information
  • Remote GPIB/USB probe control through the Oscilloscope
  • AutoZero:  Zeros out output offset
  • Applications:
    • High-speed switch mode power supply design
    • CAN/LIN bus design
    • High-speed digital design
    • Digital design and characterization
    • Manufacturing engineering and test
    • Research and development
  • Small compact probe head for probing small geometry circuit elements
  • Straight pin, square pin, solder down, variable pitch standard accessories
  • Robust design for reliability
  • Outstanding electrical performance
  • Versatile device-under-test connectivity
  • Ease of use
  • for probe setup control
  • for operating status information
  • High-speed switch mode power supply design
  • CAN/LIN bus design
  • High-speed digital design
  • Digital design and characterization
  • Manufacturing engineering and test
  • Research and development
  • Probe Bandwidth:  1 GHz
  • Probe Attenuation:  5X, 50X
  • Probe Rise Time:  ≤350 ps
  • Diffeial Input Capacitance: 
  • Diffeial Input Resistance:  1 MΩ
  • CMRR:
    • >55 dB at 30 kHz
    • >50 dB at 1 MHz
    • >18 dB at 250 MHz (warranted at 50X attenuation)
  • Sensitivity /Noise Level:
    • 2 mVRMS (5X attenuation)
    • 1 mVRMS (50X attenuation) referred to the probe output
    • 10 mVRMS (5X attenuation)
    • 50 mVRMS (50X attenuation) referred to the probe input
  • Diffeial Mode Input Voltage:  ±42 V (DC + pk AC); 30 VRMS
  • Common Mode Input Voltage:  ±35 V (DC + pk AC); 25 VRMS
  • Input Offset Range:  ±42 V (5X or 50X attenuation)
  • Max Input Voltage (Nondestruct):  ±100 V (DC + pk AC)
  • Selectable Bandwidth Filter Limits:  100 Hz, 10 kHz, 1 MHz, Full
  • DC Reject:
    • 0.4 Hz (5X)
    • 4 Hz (50X)
  • Gain Accuracy at DC:  ±2%
  • Propagation Delay:  6.5 ns
  • Probe Interface:  TekVPI™ Probe Interface
  • >55 dB at 30 kHz
  • >50 dB at 1 MHz
  • >18 dB at 250 MHz (warranted at 50X attenuation)
  • 2 mVRMS (5X attenuation)
  • 1 mVRMS (50X attenuation) referred to the probe output
  • 10 mVRMS (5X attenuation)
  • 50 mVRMS (50X attenuation) referred to the probe input
  • 0.4 Hz (5X)
  • 4 Hz (50X)
  •       

    Tektronix

    Tektronix TDP1000 Product Overview