Manufacturer Logo
Tektronix

Tektronix TIVM02

Differential Probe, 200 MHz, 5X/50X, 3M Cable
500 MHz Szerokość pasma
2 Liczba kanałów
100:1 Współczynnik tłumienia
1500 V RMS (CAT III) Maksymalne napięcie wejściowe
10 MΩ Impedancja wejściowa
Zamów kartę katalogową
Atrybut Wartość
Szerokość pasma 500 MHz
Liczba kanałów 2
Współczynnik tłumienia 100:1
Maksymalne napięcie wejściowe 1500 V RMS (CAT III)
Impedancja wejściowa 10 MΩ
Pojemność wejściowa 3,5 pF
Czas narastania 700 ps
Zakres temperatur pracy 0°C do +50°C
Kategoria bezpieczeństwa CAT III 1500 V, CAT IV 600 V
Długość kabla 1,2 m
Tektronix TIVM02 Options:
  • TIVM02 - C3 - Calibration Service 3 Years
  • TIVM02 - C5 - Calibration Service 5 Years
  • TIVM02 - D1 - Calibration Data Report
  • TIVM02 - D3 - Calibration Data Report 3 Years (with Opt. C3)
  • TIVM02 - D5 - Calibration Data Report 5 Years (with Opt. C5)
  • TIVM02 - G3 - Complete Care 3 Years (includes loaner, scheduled calibration, and more)
  • TIVM02 - R3 - Repair Service 3 Years (including warranty)
  • TIVM02 - R5 - Repair Service 5 Years (including warranty)
Formularz kalibracji Tektronix TIVM02

Formularz naprawy Tektronix TIVM02
Maximum file size: 40MB

No Related Posts

Więcej informacji
ModelTIVM02

 TEKTRONIX TIVM02 Differential Probe

 

The Tektronix TIVM02 IsoVu Measurement System offers a galvanically isolated measurement solution for accurately resolving high bandwidth, differential signals up to ±50 Vpk in the presence of large common mode voltages with the best in class common mode rejection performance across its bandwidth.  The Tektronix TIVM02 (IsoVu) products can be used on most Tektronix oscilloscopes with the TekVPI interface and on MSO/DPO70K series oscilloscopes with the TCA-VPI50 adapter. IsoVu utilizes an electro-optic sensor that converts the electrical signal from the sensor tip cables to an optical signal, which electrically isolates the device-under-test from the oscilloscope.

IsoVu incorporates four separate lasers, an optical sensor, five optical fibers, and sophisticated feedback and control techniques. The sensor head, which connects to the test point, has complete electrical isolation and is powered over one of the optical fibers (No batteries required). IsoVu is an ideal solution for users making the following measurements:

Differential measurements in the following conditions:

  • Complete galvanic isolation is required
  • High common mode voltage
  • High frequency common mode interference
  • High frequency measurements
  • Measurements in high EMI environments
  • EMI compliance testing
  • ESD testing

 

Specifications for the  TEKTRONIX TIVM02 Differential Probe Include:

Bandwidth 200 MHz
Rise Time ‰¤ 1.8 ns (typical)
Attenuation up to 50X
Common mode voltage 60 kV peak

 