Yokogawa 701251

ManufacturerYokogawa
Model701251
ProductAnalog Voltage Input Module, High-Speed 1 Ms/s 16-Bit Isolation Module (2 ch)
TypeUsed

 

Yokogawa 701250

Yokogawa 701250

Yokogawa 701260

Yokogawa 701260

Get Quote
SKU
tw701251

YOKOGAWA 701251 ANALOG VOLTAGE INPUT MODULE

The Yokogawa 701251 Analog Voltage Input Module is a two input channel, high speed, isolation module designed to provide a high sensitivity range, low noise, and high noise immunity.  

Features and Specifications of the Yokogawa 701251 Analog Voltage Input Module include:

Features:

  • High Speed:  1Ms/s
  • High Sensitivity Range:  10 mV
  • Low Noise:   ±100  Î¼V (typical)
  • High noise immunity

Specifications:

  • Effective measurement range:  Twice of setting range
  • Number of input channels:  2
  • Input coupling:  AC, DC, and GND
  • Maximum sample rate:  1 MS/s  
  • Input format:  Isolated unbalanced
  • Frequency characteristics (-3 dB point when a sine wave of amplitude 60% of range is input):  
    • 50 mV to 200 V range: DC to 300 kHz
    • 20 mV and 10 mV range: DC to 200 kHz
  • Voltage-axis range setting:  10 mV to 200 V range (1-2-5 steps)
  • Maximum input voltage (at frequency of 1 kHz or less):
    • Combined with the 700929 (10:1):  600 V (DC+ACpeak)
    • Combined with the 701901+701954  (1:1):  140 V (DC+ACpeak)
    • Direct input or cable not complying w/safety standard: 140V  (DC+ACpeak)
  • Maximum allowable common mode voltage (at a frequency of 1 kHz or less):
    • Working voltage of safety standard  
    • Combined with the 700929 (10:1) or combined with the 701901+701954 (1:1):  400 Vrms  (CAT l),  300 Vrms (CAT ll)
    • Direct input or cable not complying w/safety standard:  42 V (DC+ACpeak)(CAT l and CAT ll, 30 Vrms)
  • Vertical (voltage) axis accuracy DC accuracy:
    • 50 mV to 200V range:   ±(0.25% of  range)
    • 20 mV range:   ±(0.3% of  range)
    • 10 mV range:   ±(0.5% of  range)
  • Input connector:  BNC connector (isolated type)
  • Input impedance:  1 MΩ ± 1%, approximately 35 pF
  • -3 dB point when AC coupled low frequency attenuation point:  10 Hz or less (1 Hz or less when using the 700929)
  • Common mode rejection ratio:  80 dB (50/60 Hz) or more (typical)
  • Residual noise level (input section shorted): ±100 μ V or   ±0.1% of range, whichever is greater (typical)
  • Withstand voltage:  1500 Vrms for 1 minute (across each terminal and earth) (60 Hz)
  • Allowable transient surge voltage (instantaneous):   ±2100 Vpeak (across each input terminal and earth)
  • Insulation resistance:  500 VDC, 10 MΩ or more  (across each input terminal and earth)
  • A/D conversion resolution:  16 bit (2400 LSB/range)
  • Temperature coefficient:
    • Zero point:
      • 50 mV to 200 V range:   ±(0.02% of  range)/ ºC(typical)      
      • 20 mV range:   ±(0.05% of  range)/ ºC(typical)    
      • 10 mV range:   ±(0.10% of  range)/ ºC(typical)
    • Gain:  10 mV to 200 V range:   ±(0.02% of  range)/ ºC(typical)
  • Bandwidth limit:  
    • Select from OFF, 40 kHz, 4 kHz, and 400 Hz
    • Cut-off characteristics:  -12 dB/OCT (typical)
  • Probe attenuation setting:
    • Voltage probe:  1:1, 10:1, 100:1, 1000:1
    • Cur probe:  10 A:1 V (for the 700937/701933), 100 A: 1 V (for the 701930/701931)
  • Reference junction compensation accuracy (When the input terminal temperature is balanced:
    • K, E, J, T, L, U, N:   ±1 C
    • R, S, B, W:   ±1.5 C
    • Au7Fe:     ±1K
  • Bandwidth limit (digital filter):  Select from OFF, 30 Hz, 8 Hz, and 2 Hz
  • Input bias current:  20 nA or less
Clicky