Sorensen
Sorensen SGI60-167
DC Power Supply, 60V, 167A, 10000W
60 V Voltaje de salida máximo
167 A Corriente de salida máxima
10000 W (10 kW) Potencia máxima
Corriente continua (CC) Tipo de salida
0,02% + 2 mV Regulación de voltaje en línea

| Atributo | Valor |
|---|---|
| Voltaje de salida máximo | 60 V |
| Corriente de salida máxima | 167 A |
| Potencia máxima | 10000 W (10 kW) |
| Tipo de salida | Corriente continua (CC) |
| Regulación de voltaje en línea | 0,02% + 2 mV |
| Regulación de corriente en línea | 0,02% + 2 mA |
| Tensión de entrada (CA) | 208/240 V CA, trifásico |
| Frecuencia de entrada | 47 - 63 Hz |
| Interfaz de programación | RS-232 / GPIB (IEEE-488) opcional |
| Dimensiones (An x Al x Pr) | 482,6 mm x 132,1 mm x 711,2 mm (3U) |
Sorensen SGI60-167 Options:
- SGI60-167 - A - Control Option: Analog
- SGI60-167 - I - Control Option: Intelligent
- SGI60-167 - C - Input Voltage 187 / 242VAC, 3 Phase
- SGI60-167 - D - Input Voltage 342 / 440VAC, 3 Phase
- SGI60-167 - E - Input Voltage 396 / 528VAC, 3 Phase
- SGI60-167 - 0A - Remote Control: No Option
- SGI60-167 - 1A - Remote Control: IEEE-488.2 + RS-232C
- SGI60-167 - 1C - Remote Control: Ethernet + RS-232C
- SGI60-167 - 1D - Remote Control: Isolated Analog Control
- SGI60-167 - 1E - Remote Control: Shaft Locks (SGA series only)
- SGI60-167 - 2A - Remote Control: Combined Options 1A+1D
- SGI60-167 - 2C - Remote Control: Combined Options 1A+1E (SGA Only)
- SGI60-167 - 2G - Remote Control: Combined Options 1C+1D
- SGI60-167 - 2H - Remote Control: Combined Options 1C+1E (SGA Only)
- SGI60-167 - 2J - Remote Control: Combined Options 1D+1E (SGA Only)
- SGI60-167 - 3C - Remote Control: Combined Options 1A+1D+1E (SGA Only)
- SGI60-167 - 3G - Remote Control: Combined Options 1C+1D+1E (SGA Only)
- SGI60-167 - AA - Process Option: No Option
- SGI60-167 - AB - Process Option: Certificate of Calibration to ANSI / NCSL Z540-1 (includes Test Data)
- SGI60-167 - AJ - Modification: Front panel dust filter - factory installed - 3U unit only
- SGI60-167 - CV - Modification: 400Hz AC input @ 208 VAC (does not carry CE, CSA or UL marks) (6U only) STD on 3U
- SGI60-167 - PF - Modification: Passive power factor correction to 0.9 (Only applicable to 40V, 60V to 800V. Included in 10V-30V, 50V and 1000V.)
- SGI60-167 - R - RoHS Compliant Version
| Model | SGI60-167 |
|---|
SORENSEN SGI200-100 DC Power Supply The Sorensen SGI200-100 represents the next generation of high power programmable DC power supplies. It is designed for exceptional load transient response, low noise and the highest power density in the industry. The power density is enhanced by a stylish front air intake allowing supplies to be stacked without any required clearance between units. The SGI combines onboard intelligent controls with the outstanding power electronics common to all SG family supplies. These controls enable sophisticated sequencing, constant power mode and save/recall of instrument settings.At the heart of the Sorensen SGI200-100 is a 5 kW power module. Four modules are configured in a single chassis to deliver 20kW of power. Combinations of these chassis can then be easily paralleled to achieve power levels up to 150 kW. Paralleled units operate like one single supply providing total system cur. Looping of sequences makes the SGI200-100 ideal for repetitive testing. An impressive vacuum fluorescent graphical display in eight languages, context sensitive “soft” keys and front panel keyboard simplify programming. Specifications of the SORENSEN SGI200-100 DC Power Supply includes: See Datasheet for Full Specifications 1D - Isolated Analog Control E - 480VAC 3-phase (+/- 10%) MSRP: $26,030 | In Stock!
The Sorensen SGI250-120 is a Programmable Precision High Power DC Power Supply that is designed for exceptional load transient response, low noise and high power density which is enhanced by a stylish front air intake allowing supplies to be stacked without any required clearance between units. The SGI250-120 combines onboard intelligent controls with outstanding power electronics. These controls enable sophisticated sequencing, constant power mode and save/recall of instrument settings. Looping of sequences makes the SGI250-120 ideal for repetitive testing. An impressive vacuum fluorescent graphical display in eight languages, context sensitive “soft” keys and front panel keyboard simplify programming of the SGI250-120.
Features of the Sorensen SGI250-120 include:High Power Density: 30kW in a 6U chassis Wide Voltage Range: 0-250Vand 0-120 Amps Fast Load Transient Response: Protection from undesired voltage excursions Low RMS and P-P Noise: Suitable for the most sensitive applications Parallelable up to 150 kW: Expandable as your requirement grows Modular Design: Upgradeable for the ultimate in investment protection Sequencing: Program custom waveforms Easy-to-read Fluorescent Display: Supports English, French, German, Italian, Spanish, Chinese, Japanese, and Korean languages 16-bit Resolution: Optional IEEE-488.2 + RS-232C + Ethernet provides precise control Ethernet Control: LXI Class C compliant communication through integrated web server Direct Relay Control: Control output and sense isolation relays, along with polarity relays. (Ethernet Option Only) Hardware Trigger: Ethernet Option Only Specifications of the Sorensen SGI250-120 include:
Common:Remote Sense: Load-line loss compensation for models <= 100 V is 10% above full scale voltage total (5% per load-line), and models > 100 V is 4% above full scale voltage total (2% per load-line). Parallel Operation: Up to 5 units may be paralleled for additional cur within the power supply single-unit specifications, with exception of the DC output cur set accuracy. Additional paralleled SG units will add 0.3% inaccuracy per unit. To parallel more than 5 units,contact factory. Series Operation: Up to 2 units (see Output Float Voltage) Input: Nominal Voltage 3 phase, 3 wire + ground: 208/220 VAC (operating range 187 - 242 VAC); 380/400 VAC (operating range 342 - 440 VAC); and 440/480 VAC (operating range 396 - 528 VAC) Frequency: 47 – 63Hz (optional 400Hz @ 208VAC, does not carry CE, UL or CSA markings) Power Factor: >0.9 typical at 208/220 VAC input; >0.78 typical at 380/400 VAC input; and >0.69 typical at 440/480 VAC input Protection (typical): ½ cycle ride-though on all three phases, 3 cycle ride through on single phase; missing phase shutdown (800V model 6.4 msec on all 3 phases) Environmental: Operating Temperature: 0 to 50º C Storage Temperature: -25º C to 65º C Humidity Range: Relative humidity up to 95% non-condensing, 0º C – 50º C Altitude: Operating full power available up to 5,000 ft. (~1,500 m), derate 10% of full power for every 1,000 feet higher; non-operating to 40,000 ft. (~12,000 m) Cooling: Front and side air inlet, rear exhaust. Temperature controlled, variable speed fans. Units may be stacked without spacing. Regulatory: Certified to UL/CSA 61010 and IEC/EN 61010-1, CE Compliant, Semi-F47 Compliant Front Panel Dust Filter: 30 PPI (Pores Per Inch) - must ensure adequate airflow and / or derate max. temperature. 3U unit only. Physical: Dimensions: Width: 19.00” (48.3 cm), Depth 25.0” (63.5 cm) Height: 5-15 kW units: 3U – 5.25” rack mount (13.34 cm) 20-30 kW units: 6U – 10.5” rack mount (26.67 cm) Weight: 3U < 80 lbs. (36 kg) 6U Output: Ripple & Noise (Voltage Mode, Typical): Ripple and noise specified at full load, nominal AC input. Noise measured with 6ft. cable, 1µf at load Ripple (Cur Mode): <+/- 0.04% of full scale rms cur DC Voltage Slew Rate: 45A / ms typical - resistive load Line Regulation (with sense wires used): (±10% of nominal AC input, constant load); Voltage Mode: +/- 0.01% of full scale; and Cur Mode: +/- 0.05% of full scale Load Regulation (with sense wires used): no load to full load, nominal AC input); Voltage Mode: +/- 0.02% of full scale; and Cur Mode: +/- 0.1% of full scale Load Transient Response: Recovers within 1ms to +/-0.75% of full-scale of steadystate output for a 50% to 100% or 100% to 50% load change Efficiency: 87% typical at nominal line and max load Stability: ±0.05% of set point after 30 minute warm-up and over 8 hours at fixed line, load and temperature Temperature Coefficient: 0.02%/ C of maximum output voltage rating for voltage set point 0.03%/ C of maximum output cur rating for cur set point Output Float Voltage: Negative terminal within +/- 300 V of chassis potential. (We recommend the use of optional isolated analog Interface (IAI).) Supplies in “series” should be the same output voltage/cur. Programming Specifications of the Sorensen SGI250-120 include:
Front Panel Display:Voltage Accuracy: +/- 0.1% of full scale Cur Accuracy: +/- 0.4% of full scale Resolution: 4.0 digits Remote Analog Interface:
Voltage Accuracy: +/-0.25% of full scale for 0-5 V range, +/-0.5% of fullscale for 0-10 V range Cur Accuracy: 0.8% of full scale Remote Digital Interface: Voltage Accuracy: +/- 0.1% of full scale Cur Accuracy: +/- 0.4% of full scale Resolution: +/-0.002% of full scale OVP: Accuracy: +/- 1% of full scale Resolution: +/-0.002% of full scale User I/O: Disconnect & Polarity-reversal relay control (Only available with Ethernet Option) Read-Back / Monitoring Specifications of the Sorensen SGI250-120 include:
Front Panel Display:Voltage Accuracy: +/- 0.1% of full scale Cur Accuracy: +/- 0.4% of full scale Resolution: 4.0 digits Remote Analog Interface:
Accuracy: +/-1.0% of full scale (0 - 10V) Remote Digital Interface: Voltage Accuracy: +/- 0.1% of full scale Cur Accuracy: +/- 0.4% of full scale Resolution: +/-0.002% of full scale User I/O: Disconnect & Polarity-reversal relay control (Only available with Ethernet Option)
Features of the Sorensen SGI60-167 Programmable Precision High Power DC Power Supply include:
10kW in a 3U chassis providse high power density Wide voltage range of 0-60V and 0-167amp Fast load transient response provides protection from undesired voltage excursions Low RMS and P-P noise for very sensitive applications Parallelable up to 150 kW making it expandable as requirements increase Upgradeable modular design Sequencing allows programming custom waveforms Easy-to-read fluorescent display supports 8 languages Context sensitive soft keys Front panel keyboard Onboard intelligent controls enable sophisticated sequencing, constant power mode and save/recall of instrument settings Precise control with 16-bit Resolution, optional IEEE-488.2, RS-232C, and Ethernet Specifications of the Sorensen SGI60-167 Programmable Precision High Power DC Power Supply include: Common: Input: Output: Remote sense: Parallel operation: Series operation: Up to 2 units (see output float voltage) Terminals are provided to sense output voltage at point of load Maximum line drop 5% of rated voltage per line Up to 5 units may be paralleled for additional cur within the power supply single-unit specifications (with exception of the DC output cur set accuracy) Additional paralleled SG units will add 0.3% inaccuracy per unit Nominal voltage 3 phase, 3 wire + ground: Frequency: 47 - 63Hz Power factor: Protection (typical): 208/220 VAC (operating range 187 - 242 VAC) 380/400 VAC (operating range 342 - 440 VAC) 440/480 VAC (operating range 396 - 528 VAC) > 0.75 typical for 208/220 VAC input > 0.72 typical for 380/480 VAC input > 0.69 typical for 440/480 VAC input 1/2 cycle ride through on all three phases 3 cycle ride through on single phase Missing phase shutdown Ripple (cur mode): <+/- 0.04% of full scale rms curren DC voltage slew rate (40-800V): ≈ < 100 ms 5-95% of full scale typical - resistive load Line regulation (with sense wires used): Load regulation (with sense wires used): ± 10% of nominal AC input, constant load Voltage mode: +/- 0.01% of full scale (40-800V) Cur mode: +/- 0.05% of full scale (40-800V) no load to full load, nominal AC input Voltage mode: +/- 0.02% of full scale (40-800V) Cur mode: +/- 0.1% of full scale Load transient response: REcovers within 1 ms to +/-0.75% of full-scale of steadystate output for a 50% to 100% or 100% to 50% load change Efficiency: 87% typical at nominal line and max load Stability: 0.05% of set point after 30 minute warm-up and over 8 hours at fixed line, load Temperature coefficient: Output float voltage: Negative terminal within +/- 300V of chassis potential 0.02%/ C of maximum output voltage rating for voltage set point, typical 0.03%/ C of maximum output cur rating for cur set point, typical
Product Overview
Sorensen SGI250-120
The Sorensen SGI250-120 is a Programmable Precision High Power DC Power Supply that is designed for exceptional load transient response, low noise and high power density which is enhanced by a stylish front air intake allowing supplies to be stacked without any required clearance between units. The SGI250-120 combines onboard intelligent controls with outstanding power electronics. These controls enable sophisticated sequencing, constant power mode and save/recall of instrument settings. Looping of sequences makes the SGI250-120 ideal for repetitive testing. An impressive vacuum fluorescent graphical display in eight languages, context sensitive “soft” keys and front panel keyboard simplify programming of the SGI250-120.
Features of the Sorensen SGI250-120 include:
Common:
Front Panel Display:
Front Panel Display:
Product Overview
SORENSEN SGI60-167 POWER SUPPLY
The Sorensen SGI60-167 is a Programmable Precision High Power DC Power Supply that is designed for exceptional load transient response, low noise and high power density which is enhanced by a stylish front air intake allowing supplies to be stacked without any required clearance between units. The SGI60-167 combines onboard intelligent controls with outstanding power electronics. Looping of sequences makes the SGI60-167 ideal for repetitive testing. Features of the Sorensen SGI60-167 Programmable Precision High Power DC Power Supply include:
- Remote sense:
- Terminals are provided to sense output voltage at point of load
- Maximum line drop 5% of rated voltage per line
- Parallel operation:
- Up to 5 units may be paralleled for additional cur within the power supply single-unit specifications (with exception of the DC output cur set accuracy)
- Additional paralleled SG units will add 0.3% inaccuracy per unit
- Series operation: Up to 2 units (see output float voltage)
- Nominal voltage 3 phase, 3 wire + ground:
- 208/220 VAC (operating range 187 - 242 VAC)
- 380/400 VAC (operating range 342 - 440 VAC)
- 440/480 VAC (operating range 396 - 528 VAC)
- Frequency: 47 - 63Hz
- Power factor:
- > 0.75 typical for 208/220 VAC input
- > 0.72 typical for 380/480 VAC input
- > 0.69 typical for 440/480 VAC input
- Protection (typical):
- 1/2 cycle ride through on all three phases
- 3 cycle ride through on single phase
- Missing phase shutdown
- Ripple (cur mode): <+/- 0.04% of full scale rms curren
- DC voltage slew rate (40-800V): ≈ < 100 ms 5-95% of full scale typical - resistive load
- Line regulation (with sense wires used):
- ± 10% of nominal AC input, constant load
- Voltage mode: +/- 0.01% of full scale (40-800V)
- Cur mode: +/- 0.05% of full scale (40-800V)
- Load regulation (with sense wires used):
- no load to full load, nominal AC input
- Voltage mode: +/- 0.02% of full scale (40-800V)
- Cur mode: +/- 0.1% of full scale
- Load transient response: REcovers within 1 ms to +/-0.75% of full-scale of steadystate output for a 50% to 100% or 100% to 50% load change
- Efficiency: 87% typical at nominal line and max load
- Stability: 0.05% of set point after 30 minute warm-up and over 8 hours at fixed line, load
- Temperature coefficient:
- 0.02%/ C of maximum output voltage rating for voltage set point, typical
- 0.03%/ C of maximum output cur rating for cur set point, typical
- Output float voltage: Negative terminal within +/- 300V of chassis potential
- Terminals are provided to sense output voltage at point of load
- Maximum line drop 5% of rated voltage per line
- Up to 5 units may be paralleled for additional cur within the power supply single-unit specifications (with exception of the DC output cur set accuracy)
- Additional paralleled SG units will add 0.3% inaccuracy per unit
- 208/220 VAC (operating range 187 - 242 VAC)
- 380/400 VAC (operating range 342 - 440 VAC)
- 440/480 VAC (operating range 396 - 528 VAC)
- > 0.75 typical for 208/220 VAC input
- > 0.72 typical for 380/480 VAC input
- > 0.69 typical for 440/480 VAC input
- 1/2 cycle ride through on all three phases
- 3 cycle ride through on single phase
- Missing phase shutdown
- ± 10% of nominal AC input, constant load
- Voltage mode: +/- 0.01% of full scale (40-800V)
- Cur mode: +/- 0.05% of full scale (40-800V)
- no load to full load, nominal AC input
- Voltage mode: +/- 0.02% of full scale (40-800V)
- Cur mode: +/- 0.1% of full scale
- 0.02%/ C of maximum output voltage rating for voltage set point, typical
- 0.03%/ C of maximum output cur rating for cur set point, typical