The Keithley 2304A Power Supply provides both voltage control and power consumption monitoring for automated testing of portable, battery-operated devices. The Keithley 2304A Power Supply is optimized for testing battery-operated, wireless communication devices such as cellular phones that undergo substantial load changes for very short time intervals. The Keithley 2304A Power Supply exhibits outstanding voltage stability during pulse load changes and can simultaneously measure load currents, even if they are short pulses. In addition, the Keithley 2304A Power Supply can sink cur and, thus, take on the characteristics of a discharged, rechargeable battery for testing chargers and charger-control circuitry.
Features and Specifications of the Keithley 2304A Power Supply include:
DC Voltage Output
Output Voltage:
0 to +20VDC (for Normal Output Response)
0 to +15VDC (for Enhanced Output Response)
Output Accuracy: ±(0.05% + 10mV)
Programming Resolution: 5mV
Readback Accuracy: ±(0.05% + 10mV)
Readback Resolution: 1mV
Output Voltage Settling Time: 5ms to within stated accuracy
Load Regulation: 0.01% + 2mV
Line Regulation: 0.5mV
Stability: 0.01% + 0.5mV
Transient Response to 1000% Load Change:
Normal Mode:
Transient Recovery Time:
<50 µs to within 100mV of previous level
<100 µs to within 20mV of previous level
Enhanced Mode:
Transient Recovery Time:
<40 µs to within 100mV of previous level
<80 µs to within 20mV of previous level
Transient Voltage Drop:
<100mV, typical
<200mV, typical
Remote Sense: Automatic, 2V max drop in each lead. Add 2mV to the voltage load regulation specification for each 1V change in the negative output lead due to load cur change.
DC Current
Output Current: 5A max (not intended to be operated in parallel)
Compliance Accuracy: ±(0.16% + 5mA)
Programmed Compliance Resolution: 1.25mA
Readback Accuracy:
5A range: ±(0.2% + 1mA)
5mA range: ±(0.2% + 1 µA)
Readback Resolution:
5A range: 100 µA
5mA range: 0.1 µA
Cur Sink Capacity:
3A max (for Normal Output Response)
1A (for Enhanced Output Response)
Load Regulation: 0.01% + 1mA
Line Regulation: 0.5mA
Stability: 0.01% + 50 µA
Digital Voltmeter Input
Input Voltage Range: 0 to +20VDC
Input Impedance: 10O typical
Maximum Voltage (either input terminal) With Respect to Output Low: €“3V, +22V
Reading Accuracy: ±(0.05% + 10mV)
Reading Resolution: 1mV
DC General
Measurement Time Choices: 0.01 to 10 PLC, in 0.01PLC steps
Average Readings: 1 to 10
Reading Time 1: 31ms, typical
Pulse Cur Measurement Operation
Trigger Level: 5mA to 5A, in 5mA steps
Trigger Delay: 0 to 100ms, in 10 µs steps
Internal Trigger Delay: 25 µs
High/Low Average Mode:
Measurement Aperture Settings: 33.3 µs to 833ms, in 33.3 µs steps
Average Readings: 1 to 100
Burst Mode:
Measurement Aperture: 33.3 µs
Conversion Rate: 3600/second, typical
Number of Samples: 1 to 5000
Transfer Samples Across IEEE Bus in Binary Mode: 4800 bytes/second, typical
Long Integration Mode : 11
Measurement time: 850ms (840ms) to 60 seconds in 16.7ms (20ms) steps
General
Isolation (Low-Earth): 22VDC max
Programming: IEEE-488.2 (SCPI)
User-Definable Power-Up States: 5
Rear Panel Connector: 8-position quick disconnect terminal block for output(4), sense (2), and DVM (2)
Relay Control Jack: 2-channel, sink 150mA max, 15V max (accepts 0.173 in.Bantam-type plug (CS-1003-1))
Power Consumption: 200VA max
Input Power: 100V €“240V AC, 50 or 60Hz (auto detected at power-up)