Manufacturer Logo
Keysight (Agilent)

Keysight (Agilent) 83640A

Synthesized Sweep Signal Generator, 10 MHz - 40 GHz
10 MHz to 40 GHz Frequency Range
-20 dBm to +10 dBm (typical) Output Power Range
1 Hz Frequency Resolution
-80 dBc/Hz at 10 kHz offset (at 10 GHz) SSB Phase Noise
Less than 5 ms (typical) Frequency Switching Speed
Request datasheet
Attribute Value
Frequency Range 10 MHz to 40 GHz
Output Power Range -20 dBm to +10 dBm (typical)
Frequency Resolution 1 Hz
SSB Phase Noise -80 dBc/Hz at 10 kHz offset (at 10 GHz)
Frequency Switching Speed Less than 5 ms (typical)
Harmonic Suppression >25 dBc
Output Connector 2.4 mm (female)
Modulation Capabilities AM, FM, Phase, and Pulse Modulation
Power Level Accuracy ±1 dB (typical)
Interface GPIB (IEEE-488)
Keysight (Agilent) 83640A Options:
  • 83640A - 001 - Add Step Attenuator
  • 83640A - 002 - Internal Modulation Generator
  • 83640A - 003 - Delete Front Panel Keyboard/Display
  • 83640A - 004 - Rear Panel RF Output
  • 83640A - 006 - Fast Pulse Modulation
  • 83640A - 008 - 1 Hz Frequency Resolution
  • 83640A - 700 - Mate System Compatibility
  • 83640A - W30 - Extended Repair Service
Calibration Form Keysight (Agilent) 83640A

Repair Form Keysight (Agilent) 83640A
Maximum file size: 40MB

No Related Posts

More Information
Model83640A

The HP/Agilent 83640A synthesized sweeper is the standard of excellence for applications requiring the high performance and accuracy of a synthesized source and the speed and versatility of a sweep oscillator. It delivers excellent harmonic performance with harmonics at least 50 dB below the carrier from 2.0 to 26.5 GHz, and at least 40 dB below the carrier above 26.5 GHz. Synthesized broadband frequency coverage and precise 1 Hz frequency resolution (Opt 008) are generated by indirect synthesis, enabling the Agilent 83640A to achieve low single-sideband phase noise performance.

Other features and specifications include:

  • Scan modulation to simulate antenna scan patterns
  • Accurate fixed and swept frequency offset for two-tone measurement  
  • User flatness correction for leveling at remote points in the test system  
  • General purpose source for use in component test, receiver test and for frequency conversion  
  • Frequency range: 10 MHz to 40 GHz  
  • Sweep modes: Step, List, Ramp  
  • Maximum power: < 26.5 GHz: +10 dBm  
  • Maximum power: > 26.5 GHz: +6.0 dBm  
  • Power sweep range: -20 dBm to maximum power  
  • Phase noise (38 GHz at 10 KHz): -70 dBc  
  • Max FM deviation (locked): +/- 8 MHz

 

Other model in this series: