Keithley 2401

Low Voltage SourceMeter, 20V, 1A, 20W

Manufacturer:
Model: Keithley 2401
Description: Low Voltage SourceMeter, 20V, 1A, 20W
Data Sheet PDF: Keithley 2401

The Keithley 2401 Low Voltage SourceMeter provides precision voltage and cur sourcing and measurement capabilities (1µV – 20V and 10pA – 1A). It is both a highly stable DC power source and a true instrument-grade 5½-digit multimeter. The power source characteristics include low noise, precision, and readback. The multimeter capabilities include high repeatability and low noise. The result is a compact, single-channel, DC parametric tester.Features and Specifications of the Keithley 2401 Low Voltage SourceMeter Instrument include:

  • Five instruments in one:
    • Voltage source
    • Cur source
    • Voltage meter
    • Cur meter
    • Ohmmeter
  • Source and sink (4-quadrant) operation
  • 0.012% basic measure accuracy with 5½-digit resolution
  • 2-, 4-, and 6-wire remote V-source and measure sensing
  • 1700 readings/second at 4½ digits via GPIB
  • Standard SCPI, GPIB, RS-232, and Keithley Trigger Link interfaces
  • Keithley LabTracer 2.0 I-V curve tracing application software (download)
  • General Characteristics:
    • Load Impedance:  Stable into 20,000pF typical
    • Common Mode Voltage:  250V DC (40V DC for Model 2440)
    • Common Mode Isolation:  >109Ω,<1000pF
    • Overrange:  105% of range, source and measure
    • Max Voltage Drop Between Input/Output and Sense Terminal:  5V
    • Max Sense Lead Resistance:  1MΩ for rated accuracy
    • Sense Imput Impedance:  >1010Ω
    • Guard Offset Voltage:  <150µV, typical
    • Source Output Modes:
      • Fixed DC level
      • Memory List (mixed function)
      • Stair (linear and log)
      • Memory Buffer:  5,000 readings @ 5 digits (two 2,500 point buffers)
        • Includes selected measured value(s) and time stamp
        • Lithium battery backup (3 yr+ battery life)
    • Source Memory List:  100 points maximum
    • Programmability:  IEEE-488 (SCPI-1995.0), RS-232, 5 user-definable power-up states plus factory default and *RST
    • Digital Interface:  Interlock: Active low input
    • Power Supply:  100V to 240V rms, 50–60Hz (automatically detected at power up), 190VA
    • Cooling:  Convection
    • EMC:  Conforms to European Union Directive 89/336/EEC, EN 61326-1
  • Voltage source
  • Cur source
  • Voltage meter
  • Cur meter
  • Ohmmeter
  • Load Impedance:  Stable into 20,000pF typical
  • Common Mode Voltage:  250V DC (40V DC for Model 2440)
  • Common Mode Isolation:  >109Ω,<1000pF
  • Overrange:  105% of range, source and measure
  • Max Voltage Drop Between Input/Output and Sense Terminal:  5V
  • Max Sense Lead Resistance:  1MΩ for rated accuracy
  • Sense Imput Impedance:  >1010Ω
  • Guard Offset Voltage:  <150µV, typical
  • Source Output Modes:
    • Fixed DC level
    • Memory List (mixed function)
    • Stair (linear and log)
    • Memory Buffer:  5,000 readings @ 5 digits (two 2,500 point buffers)
      • Includes selected measured value(s) and time stamp
      • Lithium battery backup (3 yr+ battery life)
  • Source Memory List:  100 points maximum
  • Programmability:  IEEE-488 (SCPI-1995.0), RS-232, 5 user-definable power-up states plus factory default and *RST
  • Digital Interface:  Interlock: Active low input
  • Power Supply:  100V to 240V rms, 50–60Hz (automatically detected at power up), 190VA
  • Cooling:  Convection
  • EMC:  Conforms to European Union Directive 89/336/EEC, EN 61326-1
  • Fixed DC level
  • Memory List (mixed function)
  • Stair (linear and log)
  • Memory Buffer:  5,000 readings @ 5 digits (two 2,500 point buffers)
    • Includes selected measured value(s) and time stamp
    • Lithium battery backup (3 yr+ battery life)
  • Includes selected measured value(s) and time stamp
  • Lithium battery backup (3 yr+ battery life)
  •     

    Keithley

    Keithley 2401 Product Overview