Keysight (Agilent) N6781A

Source / Measurement Unit, 20 V, +/-1 A or 6 V, +/-3 A, 20 W

Manufacturer:
Model: Keysight (Agilent) N6781A
Description: Source / Measurement Unit, 20 V, +/-1 A or 6 V, +/-3 A, 20 W
Data Sheet PDF: Keysight (Agilent) N6781A

Keysight N6777A (Agilent) The Keysight N6777A (Agilent) is a 300W basic DC power module that provides programmable voltage and cur, measurement and protection features at a very economical price, making this module suitable to power the DUT or to provide power for ATE system resources such as fixture control.The Keysight N6777A DC Power Supply Module is used in both the small, multiple output 1U high N6700 Low-Profile Modular Power System (for ATE systems) and the multiple output N6705 DC Power Analyser (for bench testing). GPIB, LAN, USB and LXI Class C compliance are standard. Select from more than 30 diffe DC power modules ranging in capability from basic to high precision, and in power from 20-500 Watts.Key features of the Keysight  N6777A (Agilent) include: Output Ratings

  • Voltage: 150 V
  • Cur: 2 A
  • Power: 300 W
  • Programming Accuracy (at 23°C ±5°C)

  • Voltage: 0.1% + 150 mV
  • Cur: 0.15% + 30 mA
  • Ripple & Noise (20 Hz to 20 MHz)

  • Voltage peak-to-peak: 68 mV
  • Voltage rms: 27 mV
  • Cur rms: 6 mA
  •         Product Overview  Keysight N6781A (Agilent) (call for availability) The Keysight N6781A is a source / measure unit (SMU) designed specifically for the task of battery drain analysis. Whether the device-under-test (DUT) is and e-Book reader, MP3 player, mobile phone, or pacemaker, the N6781A’s seamless measurement ranging, programmable output resistance, and auxiliary DVM combine to be the best set of advanced features on the market for battery drain analysis.When used with the 14585A Control and Analysis software, the Keysight N6781A (Agilent) becomes an even more powerful battery drain analysis solution, offering additional insights into your measurements. Features of the Keysight N6781A (Agilent) Source / Measure Unit include:

  • Glitch-free sourcing and measurement – change output and measurement ranges without any glitches
  • 2-quadrant operation – perform as a voltage or cur source or as a CV or CC electronic load
  • Four cur measurement ranges – precisely capture the cur levels of the DUT during diffe modes of operation
  • Excellent transient response – keep voltage or cur stable during high-speed load changes
  • Fast modulation of DC – create waveforms up to 100 kHz to stimulate or load down the DUT
  • High-speed digitized measurements – capture/view the power consumption of the DUT every 5µs with built-in 200 kHz digitizer
  • Specifications of the Keysight N6781A (Agilent) Source / Measure Unit include:Output Ratings:

  • Voltage: Up to 20V
  • Cur: Up to ±3A
  • Power: 20W
  • Programming Accuracy (at 23°C ± 5°C after 30 minute warm-up. Applies from min. to max. program):

  • Voltage, 20V range: 0.025% + 1.8mV
  • Voltage, 6V range: 0.025% + 600 µV
  • Cur, 3A and 1A ranges: 0.04% + 300 µA
  • Cur, 300mA range: 0.03% + 150 µA
  • Transient Response:

  • < 30 mV of droop in response to 0.15A / µs cur transients
  • Measurement Accuracy (at 23°C ± 5°C):

  • Voltage, 20V range: 0.025% + 1 mV*
  • Cur, 3A range: 0.03% + 200 µA*
  • Cur, 100mA range: 0.025% + 7 µA*
  • Cur, 10 µA range: 0.025% + 8 nA
  •     

    Keysight Agilent