Additel

Additel ADT927-N

10,000PSI Hydraulic Test Pump
0 to 3,000 psi (0 to 200 bar) Pressure Range
±0.05% FS (Full Scale) Accuracy
Gauge Pressure Pressure Type
Gas and liquid compatible (non-corrosive media) Media Compatibility
1/4 NPT Male Process Connection
Request datasheet
Attribute Value
Pressure Range 0 to 3,000 psi (0 to 200 bar)
Accuracy ±0.05% FS (Full Scale)
Pressure Type Gauge Pressure
Media Compatibility Gas and liquid compatible (non-corrosive media)
Process Connection 1/4 NPT Male
Output Signal 4-20 mA analog output
Power Supply 9 to 32 VDC
Operating Temperature -10°C to 70°C (14°F to 158°F)
Ingress Protection IP65
Wetted Materials 316L Stainless Steel
Additel ADT927-N Options:
  • ADT927-N - N - 1/4NPT female pressure port
  • ADT927-N - N2 - 1/2NPT female pressure port
  • ADT927-N - B - 1/4BSP female pressure port
  • ADT927-N - B2 - 1/2BSP female pressure port
  • ADT927-N - M - M20X1.5 female pressure port
  • ADT927-N - SET - Pump and carrying case included
  • ADT927-N - W - Water media (deionized water) instead of default oil media
Calibration Form Additel ADT927-N

Repair Form Additel ADT927-N
Maximum file size: 40MB

No Related Posts

More Information
ModelADT927-N

 Additel  ADT927-N  10,000PSI Hydraulic Test Pump

 

The  Additel ADT927-N Hydraulic Test Pump  is a hand operated pressure pump designed to generate pressure from 85% vacuum to 10,000 psi (700 bar). With the patented screw press technology, the high pressure can be easily generated, as well as increased and decreased smoothly.  

Additel ADT927-N Hydraulic Test Pump  is an ideal comparison test pump for calibrating pressure measuring instruments such as test gauges, indicators or transducers in the field or laboratory. With no check valve (non-returning valve), the ADT927-N avoids the troublesome leakage issues that is usually experienced with most hand pumps and allows for minimal maintenance. Two hand-tight quick connectors installed on the Additel ADT927-N Hydraulic press Test Pump allow easy connecting and disconnecting to the test pump without the need for PTFE tape or wrenches.

 

Specifications for the  Additel ADT927-N 10,000PSI Hydraulic Test Pump Include:

Generated Pressure Range 85% vacuum to 10,000psi (700bar)
Reservoir capacity 245 ml
Media Oil or deionized water
Piston volume 19 ml (Low pressure priston),  0.9 ml (High pressure priston)

 