emtest
emtest HFK
Capacitive Coupling Clamp
IEC 61000-4-12, IEC 61000-4-18 Test Standard Compliance
Damped Oscillatory Wave (Ring Wave) Output Waveform
Up to 4 kV (open circuit) Peak Voltage Output
Up to 2 kA (short circuit) Peak Current Output
100 kHz and 1 MHz oscillatory waveforms Frequency Range

| Attribute | Value |
|---|---|
| Test Standard Compliance | IEC 61000-4-12, IEC 61000-4-18 |
| Output Waveform | Damped Oscillatory Wave (Ring Wave) |
| Peak Voltage Output | Up to 4 kV (open circuit) |
| Peak Current Output | Up to 2 kA (short circuit) |
| Frequency Range | 100 kHz and 1 MHz oscillatory waveforms |
| Coupling/Decoupling Network | Integrated CDN for AC and DC power lines |
| Repetition Rate | Adjustable, up to 40 pulses per second |
| Polarity | Positive and Negative, selectable |
| Control Interface | PC-controlled via software interface (TEMA) |
| Power Supply Input | 100–240 V AC, 50/60 Hz |
emtest HFK Options:
- HFK - CA EFT Kit - Calibration Kit for EFT Burst calibration includes: KW 50 50ohm load resistor, KW 1000 1,000ohm load resistor, AD-CA EFT Adapter 4mm/6mm to coaxial F103 connector (connection Load resistor to transducer plate)
- HFK - AD-CA EFT - Adapter to match the transducer plate to the KW 50 load resistor (connection Load resistor to transducer plate)
- HFK - KW 50 - Coaxial 50ohm load resistor for EFT/Burst transient calibration
| Model | HFK |
|---|
emtest HFK Capacitive Coupling ClampThe emtest HFK is used to couple EFT/burst pulses to I/O lines as required in diffe European and international standards for immunity testing. The coupling of the Electrical Fast Transients EFT/burst to signal lines can usually not be achieved by discrete capacitive coupling without interfering with the signal flow. It is often impossible to contact the required circuit (direct), e. g. coaxial or shielded cables. In this case the coupling is realized by the capacitive coupling clamp. The interference simulator can be connected on both sides of the coupling clamp.
Features of the emtest HFK include:Connects directly to the emtest UCS 500N5. And with adapters, it connects to the emtest NX5 Compact Immunity Tester Construction as per IEC 61000-4-4 Ed.3 EFT/Burst testing of signal- and datalines Active coupling length 1m Permissable burst voltage 7 kV For cable diameter up to 40 mm The HFK Capactive Coupling Clamp is identical to the emtest CCI Capacitive Coupling Clamp, except that it requires adapters to connect the NX5 without the need for adapters.
Features of the emtest HFK include: