Test Equity

Test Equity 115

Temperature Chamber, -73C to +175C, 1.55 Cu Ft
-15°C a +35°C Rango de Temperatura
±0,5°C Precisión de Temperatura
±0,3°C Uniformidad de Temperatura
1,5 pies cúbicos (42,5 litros) Volumen Interior
35,6 cm x 35,6 cm x 33,0 cm Dimensiones Interiores (An x Al x Pr)
Solicitar ficha técnica
Atributo Valor
Rango de Temperatura -15°C a +35°C
Precisión de Temperatura ±0,5°C
Uniformidad de Temperatura ±0,3°C
Volumen Interior 1,5 pies cúbicos (42,5 litros)
Dimensiones Interiores (An x Al x Pr) 35,6 cm x 35,6 cm x 33,0 cm
Dimensiones Exteriores (An x Al x Pr) 63,5 cm x 68,6 cm x 63,5 cm
Alimentación Eléctrica 120 V CA, 60 Hz
Consumo de Potencia 750 W
Controlador Controlador digital de microprocesador con pantalla LCD
Certificaciones UL, CE
Test Equity 115 Options:
  • 115 - GPIB - GPIB Interface
  • 115 - SHELF - Additional Shelves
  • 115 - PORT - Additional Ports
  • 115 - 50HZ - 50 Hz Export Version
Formulario de calibración Test Equity 115

Formulario de reparación Test Equity 115
Maximum file size: 40MB

No Related Posts

Más información
Model115

 TEST EQUITY 115 Temperature Chamber

 

The TestEquity Model 115 provides full-range temperature testing in a compact floor-standing or benchtop package. Its rugged design includes a variety of standard features, making it an exceptional value.  Up to 256 steps can be programmed into as many as 40 nameable profiles. The context sensitive information key and guided steps make profile programming fast and easy. A fourline backlit LCD displays programming, setup, operating and help information.   A large LED readout indicates the actual chamber temperature with 0.1 ° resolution.

Non-CFC refrigerants are used in the Test Equity 115 in a cascade (two compressors) configuration. Pressure controls keep the operating pressures within safe limits regardless of load conditions. Liquid injection ensures cool compressor operation during high-temperature cool down for long life.  The nichrome air-heater has a low watt-density and even heat distribution for reliable performance and fast response. A fusible link provides fail-safe protection against thermal runaways, in addition to the microprocessor-based high/low limit controller.   The side and rear panels are removable to permit complete access to the refrigeration system for ease of service and preventive maintenance.

 

Specifications for the  TEST EQUITY 115 Temperature Chamber Include:

Temperature Range -73 °C to +175 °C
Internal Dimensions 16" W x 12" H x 14" D (1.55 Cubic Feet)
Heat Up Time 5 °C/minute typical
Control Tolerance ±0.2 °C

 