Fluke 5220A

| Attribute | Value |
|---|---|
| Function | Transconductance Amplifier |
| Output Current Range | Up to 220 A AC/DC |
| Input Voltage | 0 to 2.2 V (from a Fluke 5500A or 5520A calibrator) |
| Transconductance | 100 A/V (nominal) |
| Frequency Range | DC, 40 Hz to 1 kHz |
| Current Accuracy (DC) | ±0.1% of output |
| Current Accuracy (AC) | ±0.15% of output (40 Hz to 1 kHz) |
| Power Requirements | 100 V, 120 V, 220 V, or 240 V AC, 50/60 Hz |
| Operating Temperature | 18°C to 28°C (64°F to 82°F) |
| Compatible Calibrators | Fluke 5500A, 5520A Multi-Product Calibrators |
Fluke 5220A Options:
- 5220A - Y5020 - Current Shunt
- 5220A - Y5000 - Interface/Buffer
- 5220A - Y5002 - Cable (Y5000 to 5220A)
- 5220A - Y5702 - Cable (5700A to 5220A)
- 5220A - M07-594 - 7" Rack Mount Kit
- 5220A - M00-270-610 - 20" Slides for Rack Mount Kit
- 5220A - M00-280-610 - 24" Slides for Rack Mount Kit
| Model | 5220A |
|---|
The Fluke 5220A Transconductance Amplifier is designed for verifying current-measuring instruments with high accuracy. It accepts input voltages from 1 to 20 V via banana jacks and delivers precise currents up to 20 A, both AC and DC. Its capabilities make it suitable for calibrating meters, shunts, or the current function of digital multimeters, ensuring reliable performance in demanding environments.
Operating as a standalone unit or in combination with the Fluke 5100 Series Calibrator, the {product_name} integrates remote interface options for seamless use in automated calibration systems. This versatility makes it ideal for laboratory and field calibration setups.
Testwall can supply the used Fluke 5220A Transconductance Amplifier, fully tested. With expert support and global delivery, Testwall ensures high-performance instrumentation with unmatched reliability and service quality. Please contact us directly if you need calibration or repair of the Fluke 5220A Transconductance Amplifier.